Your browser doesn't support javascript.
loading
Clinical validation of two volumetric absorptive microsampling devices to support home-based therapeutic drug monitoring of immunosuppression.
Leino, Abbie D; Takyi-Williams, John; Park, Jeong M; Norman, Silas P; Sun, Duxin; Farris, Karen B; Pai, Manjunath P.
Afiliação
  • Leino AD; Department of Clinical Pharmacy, College of Pharmacy, University of Michigan, Ann Arbor, MI, USA.
  • Takyi-Williams J; Pharmacokinetic & Mass Spectrometry Core, College of Pharmacy, University of Michigan, Ann Arbor, MI, USA.
  • Park JM; Department of Clinical Pharmacy, College of Pharmacy, University of Michigan, Ann Arbor, MI, USA.
  • Norman SP; Division of Nephrology, Department of Internal Medicine, University of Michigan, Ann Arbor, MI, USA.
  • Sun D; Pharmacokinetic & Mass Spectrometry Core, College of Pharmacy, University of Michigan, Ann Arbor, MI, USA.
  • Farris KB; Department of Pharmaceutical Sciences, College of Pharmacy, University of Michigan, Ann Arbor, MI, USA.
  • Pai MP; Department of Clinical Pharmacy, College of Pharmacy, University of Michigan, Ann Arbor, MI, USA.
Br J Clin Pharmacol ; 90(11): 2897-2909, 2024 Nov.
Article em En | MEDLINE | ID: mdl-39051148
ABSTRACT

AIMS:

Dried blood volumetric absorptive microsamples (VAMS) may facilitate home-based sampling to enhance therapeutic drug monitoring after transplantation. This study aimed to clinically validate a liquid chromatography-tandem mass spectrometry assay using 2 VAMS devices with different sampling locations (Tasso-M20 for the upper arm and Mitra for the finger). Patient preferences were also evaluated.

METHODS:

Clinical validation was performed for tacrolimus and mycophenolic acid by comparison of paired VAMS and venipuncture samples using Passing-Bablok regression and Bland-Altman analysis. Conversion of mycophenolic acid VAMS to serum concentrations was evaluated using haematocrit-dependent formulas and fixed correction factors defined a priori. Patients' perspectives, including useability, acceptability and feasibility, were also investigated using established questionnaires.

RESULTS:

Paired samples (n = 50) were collected from 25 kidney transplant recipients. Differences for tacrolimus whole-blood concentration were within ±20% for 86 and 88% of samples from the upper arm and fingerstick, respectively. Using correction factors of 1.3 for the upper-arm and 1.47 for finger-prick samples, 84 and 76% of the paired samples, respectively, were within ±20% for mycophenolic acid serum concentration. Patient experience surveys demonstrated limited pain and acceptable useability of the upper-arm device.

CONCLUSIONS:

Tacrolimus and mycophenolic acid can be measured using 2 common VAMS devices with similar analytical performance. Patients are supportive of home-based monitoring with a preference for the Tasso-M20 device.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Transplante de Rim / Tacrolimo / Monitoramento de Medicamentos / Espectrometria de Massas em Tandem / Teste em Amostras de Sangue Seco / Imunossupressores / Ácido Micofenólico Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Br J Clin Pharmacol Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Transplante de Rim / Tacrolimo / Monitoramento de Medicamentos / Espectrometria de Massas em Tandem / Teste em Amostras de Sangue Seco / Imunossupressores / Ácido Micofenólico Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Br J Clin Pharmacol Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos
...