Accuracy of intraocular lens calculation formulas based on swept-source OCT biometer in cataract patients with phakic intraocular lens.
BMC Ophthalmol
; 24(1): 326, 2024 Aug 05.
Article
em En
| MEDLINE
| ID: mdl-39103785
ABSTRACT
PURPOSE:
To research the accuracy of intraocular lens (IOL) calculation formulas and investigate the effect of anterior chamber depth (ACD) and lens thickness (LT) measured by swept-source optical coherence tomography biometer (IOLMaster 700) in patients with posterior chamber phakic IOL (PC-pIOL).METHODS:
Retrospective case series. The IOLMaster 700 biometer was used to measure axial length (AL) and anterior segment parameters. The traditional formulas (SRK/T, Holladay 1 and Haigis) with or without Wang-Koch (WK) AL adjustment, and new-generation formulas (Barret Universal II [BUII], Emmetropia Verifying Optical [EVO] v2.0, Kane, Pearl-DGS) were utilized in IOL power calculation.RESULTS:
This study enrolled 24 eyes of 24 patients undergoing combined PC-pIOL removal and cataract surgery at Xiamen Eye Center of Xiamen University, Xiamen, Fujian, China. The median absolute prediction error in ascending order was EVO 2.0 (0.33), Kane (0.35), SRK/T-WKmodified (0.42), Holladay 1-WKmodified (0.44), Haigis-WKC1 (0.46), Pearl-DGS (0.47), BUII (0.58), Haigis (0.75), SRK/T (0.79), and Holladay 1 (1.32). The root-mean-square absolute error in ascending order was Haigis-WKC1 (0.591), Holladay 1-WKmodified (0.622), SRK/T-WKmodified (0.623), EVO (0.673), Kane (0.678), Pearl-DGS (0.753), BUII (0.863), Haigis (1.061), SRK/T (1.188), and Holladay 1 (1.513). A detailed analysis of ACD and LT measurement error revealed negligible impact on refractive outcomes in BUII and EVO 2.0 when these parameters were incorporated or omitted in the formula calculation.CONCLUSION:
The Kane, EVO 2.0, and traditional formulas with WK AL adjustment displayed high prediction accuracy. Furthermore, the ACD and LT measurement error does not exert a significant influence on the accuracy of IOL power calculation formulas in highly myopic eyes implanted with PC-pIOL.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Refração Ocular
/
Catarata
/
Biometria
/
Tomografia de Coerência Óptica
/
Lentes Intraoculares Fácicas
Limite:
Adult
/
Aged
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
BMC Ophthalmol
Assunto da revista:
OFTALMOLOGIA
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
China