Results 1 - 20 de 26
Filter
1.
J Nucl Med Technol ; 52(3): 256-260, 2024 Sep 05.
Article in English | MEDLINE | ID: mdl-39019573

ABSTRACT

A new 90Y SIR-Spheres delivery kit (SIROS D-vial and shield) has been introduced with a different physical form from the legacy V-Vial kit. Here, we establish the dose calibrator settings and exposure-rate-to-activity conversion factor to assay 90Y SIR-Spheres activity in the new SIROS kit. Methods: Eight D-vials with initial 90Y activities from 1.2 to 6.6 GBq within acrylic shields were assayed with dose calibrators and exposure-rate meters until activities decayed to approximately 0.1 GBq. The dose calibrator settings resulting in the lowest median activity errors and the best-fit slope of exposure rate versus activity were identified. Results: SIROS D-vial 90Y activity can be accurately and reliably estimated directly using setting 51 × 10 on both the CRC-15R and the CRC-55tR dose calibrators (errors within ±0.5%) and indirectly with an exposure-rate reading at 30 cm using conversion factor 0.664 ± 0.003 GBq/(mR/h) (R 2 = 0.985). Conclusion: Dose calibrator settings and exposure-rate-to-activity conversion factor for 90Y activity assays with new SIROS kit should be updated from legacy V-Vial parameters to avoid an approximately 10% underestimation.


Subject(s)
Yttrium Radioisotopes , Calibration , Radiometry/instrumentation , Radiometry/methods , Microspheres
2.
Appl Radiat Isot ; 209: 111326, 2024 Jul.
Article in English | MEDLINE | ID: mdl-38701595

ABSTRACT

Thorium-227 was separated from its progeny and standardized for activity by the triple-to-double coincidence ratio (TDCR) method of liquid scintillation counting. Confirmatory liquid scintillation-based measurements were made using efficiency tracing with 3H and live-timed anticoincidence counting (LTAC). The separation time and the efficiency of the separation were confirmed by gamma-ray spectrometry. Calibrations for reentrant pressurized ionization chambers, including commercial radionuclide calibrators, and a well-type NaI(Tl) detector are discussed.

3.
Radiol Phys Technol ; 17(2): 396-401, 2024 Jun.
Article in English | MEDLINE | ID: mdl-38517669

ABSTRACT

This study investigates the feasibility of estimating the radioactivity of radiopharmaceuticals using shielded syringes. The radioactivities of 99mTc-MDP, 99mTc-HMDP, 99mTc-ECD, 99mTc-MAG3, and 123I-IMP were measured using a dose calibrator. Correlation coefficients and regression equations were obtained from the radioactivity in the shielded and unshielded syringes. 99mTc-MDP was also measured for residual radioactivity after the administration. The correlation coefficients of 99mTc-MDP, 99mTc-HMDP, 99mTc-ECD, 99mTc-MAG3, and 123I-IMP were rs = 0.9998, rs = 0.9997, rs = 0.9999, rs = 0.9998, and rs = 0.9888, respectively. The regression equations were y = 0.0364x + 0.0913, y = 0.0349x + 0.0273, y = 0.0343x - 0.0018, y = 0.0522x + 0.1215, and y = 0.0383x + 0.0058, respectively. The correlation coefficient for the residual radioactivity of 99mTc-MDP was rs = 0.9887 and the regression equation was y = 0.1505x + 0.0853. The radioactivity of 99mTc- and 123I-labeled radiopharmaceuticals in shielded syringes was accurately measured. It was suggested that the measuring shielded syringes could provide an estimate of the actual radioactivity.


Subject(s)
Feasibility Studies , Iodine Radioisotopes , Radiopharmaceuticals , Syringes , Technetium , Radiopharmaceuticals/chemistry , Technetium/chemistry , Radioactivity , Calibration
4.
Molecules ; 28(3)2023 Jan 31.
Article in English | MEDLINE | ID: mdl-36771012

ABSTRACT

44Sc is a promising radionuclide for positron emission tomography (PET) in nuclear medicine. As a part of the implementation of a production site for 44Sc, precise knowledge of the activity of the product is necessary. At the Paul Scherrer Institute (PSI) and the University of Bern (UniBE), 44Sc is produced by enriched 44CaO-target irradiation with a cyclotron. The two sites use different techniques for activity measurement, namely a dose calibrator at the PSI and a gamma-ray spectrometry system at UniBE and PSI. In this work, the 44Sc was produced at the PSI, and samples of the product were prepared in dedicated containers for onsite measurements at PSI, UniBE, and the Institute of Radiation Physics (IRA) in Lausanne for precise activity measurement using primary techniques and for the calibration of the reference ionization chambers. An accuracy of 1% was obtained for the activity measurement, allowing for a precise calibration of the dose calibrator and gamma-ray spectrometry of the two production sites. Each production site now has the capability of measuring 44Sc activity with an accuracy of 2%.

5.
Appl Radiat Isot ; 190: 110473, 2022 Dec.
Article in English | MEDLINE | ID: mdl-36201936

ABSTRACT

An activity standard for 212Pb in equilibrium with its progeny was realized, based on triple-to-double coincidence ratio (TDCR) liquid scintillation (LS) counting. A Monte Carlo-based approach to estimating uncertainties due to nuclear decay data (branching ratios, beta endpoint energies, γ-ray energies, and conversion coefficients for 212Pb and 208Tl) led to combined standard uncertainties ≤ 0.20 %. Confirmatory primary measurements were made by LS efficiency tracing with tritium and 4παß(LS)-γ(NaI(Tl)) anticoincidence counting. The standard is discussed in relation to current approaches to 212Pb activity calibration. In particular, potential biases encountered when using inappropriate radionuclide calibrator settings are discussed.


Subject(s)
Lead , Scintillation Counting , Scintillation Counting/methods , Radioisotopes , Tritium/analysis , Reference Standards
6.
EJNMMI Phys ; 9(1): 19, 2022 Mar 14.
Article in English | MEDLINE | ID: mdl-35286498

ABSTRACT

BACKGROUND: 161Tb draws an increasing interest in nuclear medicine for therapeutic applications. More than 99% of the emitted gamma and X-rays of 161Tb have an energy below 100 keV. Consequently, precise activity measurement of 161Tb becomes inaccurate with radionuclide dose calibrators when using inappropriate containers or calibration factors to account for the attenuation of this low energy radiation. To evaluate the ionization chamber response, the sample activity must be well known. This can be performed using standards traceable to the Système International de Référence, which is briefly described as well as the method to standardize the radionuclides. METHODS: In this study, the response of an ionization chamber using different container types and volumes was assessed using 161Tb. The containers were filled with a standardized activity solution of 161Tb and measured with a dedicated ionization chamber, providing an accurate response. The results were compared with standardized solutions of high-energy gamma-emitting radionuclides such as 137Cs, 60Co, 133Ba and 57Co. RESULTS: For the glass vial type with an irregular glass thickness, the 161Tb measurements gave a deviation of 4.5% between two vials of the same type. The other glass vial types have a much more regular thickness and no discrepancy was observed in the response of the ionization chamber for these type of vials. Measurements with a plastic Eppendorf tube showed stable response, with greater sensitivity than the glass vials. CONCLUSION: Ionization chamber measurements for low-energy gamma emitters (< 100 keV), show deviation depending on the container type used. Therefore, a careful selection of the container type must be done for activity assessment of 161Tb using radionuclide dose calibrators. In conclusion, it was highlighted that appropriate calibration factors must be used for each container geometry when measuring 161Tb and, more generally, for low-energy gamma emitters.

7.
Appl Radiat Isot ; 184: 110161, 2022 Jun.
Article in English | MEDLINE | ID: mdl-35248797

ABSTRACT

Interest in targeted cancer therapy with alpha-emitting radionuclides is growing. To evaluate emerging radiotherapeutic agents requires precise activity measurements for consistent dose-response relationships and patient-specific dosimetry. National metrology institutes around the world have reported on the development and comparison of activity standards for medically important alpha emitters. This review describes the relevant methods and models underpinning these standards, the generation of new nuclear decay data, and the impacts on preclinical and clinical activity assays using radionuclide calibrators and γ-ray spectrometry.


Subject(s)
Radioisotopes , Radiometry , Calibration , Humans , Radioisotopes/therapeutic use , Radiometry/methods , Reference Standards , Spectrometry, Gamma
8.
J Nucl Med Technol ; 49(2): 138-142, 2021 Jun.
Article in English | MEDLINE | ID: mdl-33419954

ABSTRACT

This multicenter study aimed to determine the reproducibility of quantitative SPECT images reconstructed using a commercially available method of ordered-subset conjugate-gradient minimization. Methods: A common cylindric phantom containing a 100 kBq/mL concentration of 99mTc-pertechnetate solution in a volume of 7 L was scanned under standard imaging conditions at 6 institutions using the local clinical protocol of each. Interinstitutional variation among the quantitative SPECT images was evaluated using the coefficient of variation. Dose calibrator accuracy was also investigated by measuring the same lot of commercially available 99mTc vials at each institution. Results: The respective radioactivity concentrations under standard and clinical conditions ranged from 95.71 ± 0.60 (mean ± SD) to 108.35 ± 0.36 kBq/mL and from 96.78 ± 0.64 to 108.49 ± 0.11 kBq/mL, respectively. Interinstitutional variation in radioactivity concentration was 4.20%. The bias in the radioactivity concentrations in SPECT images was associated with the accuracy of the dose calibrator at each institution. Conclusion: The reproducibility of the commercially available quantitative SPECT reconstruction method is high and comparable to that of PET, for comparatively large (∼7 L), homogeneous objects.


Subject(s)
Algorithms , Tomography, Emission-Computed, Single-Photon , Phantoms, Imaging , Reproducibility of Results , Technetium
9.
Curr Radiopharm ; 13(1): 80-85, 2020.
Article in English | MEDLINE | ID: mdl-31258094

ABSTRACT

BACKGROUND: 99mTc is a radioactive isotope that is obtained by eluting a 99Mo/99mTc generator. (PINSTECH, Islamabad) and used for radionuclide scanning. OBJECTIVES: The objective of this work is to study the uncertainties in 99mTc activity that exist due to time delay between injection preparation and administration to patients, during the process of gamma camera scanning. METHODS: Lead canisters were used for storing elution vials and dose calibrator for measuring 99mTc activity in mCi. The activity of preparing 99mTc injection and its administration to patients were compared with the prescribed values of activity recommended in the Society of Nuclear Medicine procedure guidelines. RESULTS: This study showed that uncertainty in the activity existed in one thyroid patient, 38 bone patients, 5 renal patients and 45 cardiac patients. CONCLUSION: This uncertainty in activity exists due to time delay between injection preparation and administration to patients, as well as due to residual radionuclide that is not injected into patients and remains in the syringe.


Subject(s)
Bone and Bones/diagnostic imaging , Heart/diagnostic imaging , Kidney/diagnostic imaging , Radiopharmaceuticals , Technetium , Thyroid Gland/diagnostic imaging , Humans , Injections , Nuclear Medicine , Radionuclide Generators , Radionuclide Imaging
10.
Appl Radiat Isot ; 153: 108829, 2019 Nov.
Article in English | MEDLINE | ID: mdl-31394471

ABSTRACT

Radionuclide activity meters depend on "calibration setting numbers" that are specific for a certain combination of radionuclide, volume and geometry. This work describes their determination for two PET emerging nuclides (44Sc and 89Zr), three commercial activity meters (CAPINTEC CRC15 BETA, ATOMLAB BIODEX 500 and VEENSTRA VDC 405) and two geometries: 1 mL plastic syringes and 5 mL glass vials. For 44Sc, only values for the dial setting for 5 mL ampoules were reported previously and its use for the assayed geometries would lead to significant differences. For 89Zr more data were available in the literature, and results are provided for new geometries. All results rely on the absolute standardisation in activity of solutions of both radionuclides.

11.
Appl Radiat Isot ; 154: 108849, 2019 Dec.
Article in English | MEDLINE | ID: mdl-31465950

ABSTRACT

A solution of 124I was standardized for activity by 4πß(LS)-γ(NaI) live-timed anticoincidence (LTAC) counting, with confirmatory measurements by triple-to-double coincidence ratio (TDCR) and CIEMAT-NIST efficiency tracing (CNET) liquid scintillation counting. The LTAC-based standard was shown to be in agreement (within k = 1 uncertainties) with previous measurements at NIST and elsewhere. Calibration settings for radionuclide calibrators were determined and a discrepancy with literature values, partially due to a calibration methodology dependent upon an erroneous setting for 18F, was identified and explained.

12.
Braz. arch. biol. technol ; Braz. arch. biol. technol;62: e19180545, 2019. tab, graf
Article in English | LILACS | ID: biblio-1019543

ABSTRACT

Abstract To present optimized chromatographic systems for radiochemical purity (RCP) evaluation of 99mTc-eluate and 99mTc-radiopharmaceuticals, as well as to assess doses calibrator reliability for routine purposes in hospital radiopharmacies. RCP was determined by different systems and radioactivity was quantified by TLC-scanner, doses calibrator and gamma-counter. Suitable and optimized systems were presented for RCP analyses. No significant differences were observed between radioactivity counting devices and, thus, doses calibrator showed reliability for RCP determination in hospital radiopharmacies.


Subject(s)
Radiochemistry/methods , Radiopharmaceuticals/standards , Chromatography/methods , Radiation Dosimeters
13.
Article in Japanese | MEDLINE | ID: mdl-29925748

ABSTRACT

PURPOSE: The AI-300 automated infusion device (Sumitomo Heavy Industries, Ltd., Tokyo, Japan) is subject to administration error as a function of smaller volumes of 18F-FDG dispensed via a three-way cock supplied with a disposable kit. The present study aimed to validate the administration accuracy of the AI-300 using an improved disposable kit for quantitative positron emission tomography (PET) assessment. METHODS: We determined administration accuracy between the improved and previous disposable kits by measuring variations in dispensed volumes and radioactive concentrations of 18F-FDG according to the criteria of the Japanese Society of Nuclear Medicine. A reference value was generated by measuring radioactivity using a standard dose calibrator. RESULTS: The values obtained using the previous kit deviated from the reference values by a maximum of -10.6%, and the deviation depended on dispensed volumes of 18F-FDG<0.25 mL. In contrast, the values were relatively stable when using the improved kit with dispensed 18F-FDG volumes < 0.25 mL. Variations in radioactive concentrations were relatively stable using the improved kit, whereas that of the previous kit was slightly unstable at high radioactive concentrations. CONCLUSION: The administration accuracy of the AI-300 using the previous kit varied considerably according to smaller dispensed volumes, but the improved kit might alleviate this problem. The present results indicated that the improved disposal kit should be immediately implemented to eliminate uncertainty surrounding quantitative PET findings.


Subject(s)
Positron-Emission Tomography , Radiopharmaceuticals , Fluorodeoxyglucose F18 , Japan , Tokyo
14.
Appl Radiat Isot ; 134: 89-94, 2018 Apr.
Article in English | MEDLINE | ID: mdl-29066214

ABSTRACT

Possible extended utilization of a 68Ge/68Ga source as a surrogate 18F reference source was studied. Some effects and uncertainties might arise from the difference of constructions between chambers that were used in the source calibration and those practically in use. To extend utilization of the surrogate source, response ratio of real 18F to 68Ge/68Ga sources were determined for three models of chambers. Monte-Carlo simulation using EGS5 code was also carried out to evaluate uncertainties expected. The present result shows that such a surrogate source would adapt to calibration even for chambers have with different construction from the reference chamber used as the determination of 18F equivalent activity of the 68Ge/68Ga surrogate source.

15.
Kaku Igaku ; 54(1): 545-549, 2017.
Article in Japanese | MEDLINE | ID: mdl-28804109

ABSTRACT

Low-energy characteristic x-rays emitted by 111In and 123I sources are easily absorbed by the containers of the sources, affecting radioactivity measurements using a dose calibrator. We examined the effects of different containers on the estimated activities. The radioactivities of 111In, 123I, 201Tl, and 99mTc were measured in containers frequently employed in clinical practice in Japan. The 111In measurements were performed in the vials A and B of the 111In-pentetreotide preparation kit and in the plastic syringe. The activities of 123I-metaiodobenzylguanidine and 201Tl chloride were measured in the prefilled glass syringes and plastic syringes. The milking vial, vial A, vial B, and plastic syringe were used to assay 99mTc. For 111In and 123I, measurements were performed with and without a copper filter. The filter was inserted into the well of the dose calibrator to absorb low-energy x-rays. The relative estimate was defined as the ratio of the activity estimated with the dose calibrator to the standard activity. The estimated activities varied greatly depending on the container when 111In and 123I sources were assayed without the copper filter. The relative estimates of 111In were 0.908, 1.072, and 1.373 in the vial A, vial B, and plastic syringe, respectively. The relative estimates of 123I were 1.052 and 1.352 in the glass syringe and plastic syringe, respectively. Use of the copper filter eliminated the container-dependence in 111In and 123I measurements. Container-dependence was demonstrated in neither 201Tl nor 99mTc measurements. The activities of 111In and 123I estimated with a dose calibrator differ greatly among the containers. Accurate estimation may be attained using the container-specific correction factor or using the copper filter.


Subject(s)
Indium Radioisotopes , Iodine Radioisotopes , Radioactivity , Radiopharmaceuticals , Calibration , Japan , Radioisotopes , Reference Standards , Syringes , Thallium Radioisotopes
16.
J Nucl Med Technol ; 45(2): 91-95, 2017 Jun.
Article in English | MEDLINE | ID: mdl-28280125

ABSTRACT

The administration accuracy and precision of an automated infusion device for positron-emitting radiotracers are directly associated with bias and variance in the SUVs of 18F-FDG PET/CT. Therefore, the accuracy of such devices must be confirmed and calibrated at locations in which they are used. The present study aimed to validate the administration accuracy of 3 automated infusion devices for quantitative PET assessment. Methods: Temporal variations as well as variations in radioactive concentrations and dispensed volumes of 18F-FDG were determined for the M-130, AI-300, and UG-05 automated infusion devices. The total-test dispensed volumes were 25, 20, and 18.5 mL, respectively. A reference value was generated by measuring amounts of radioactivity using a standard dose calibrator. Administration accuracy was validated according to the criteria of the Japanese Society of Nuclear Medicine. Results: The temporal variation in the M-130 and UG-05 for a specified 185 MBq was relatively stable, in the range of -1.60%-0.92% and 1.16%-5.35%, respectively, whereas that in the AI-300 was -0.55%-8.68%. For the M-130 and UG-05 devices, the difference between measured and reference value was in the range of -5%-5%. The values measured by the AI-300 deviated from the reference values by a maximum of 30%, which depends on radioactive concentration and dispensed volume of 18F-FDG. Conclusion: The administration accuracy of the AI-300 varied considerably under different conditions, but a software update might somewhat improve this. Our findings indicate that dispensed volumes of 18F-FDG should be carefully considered when the radioactive concentration is high. Administration accuracy should be regularly confirmed at each location to maintain the quality of quantitative PET assessment. The present study provides useful information about how to confirm the administration accuracy of automated infusion devices.


Subject(s)
Positron-Emission Tomography/instrumentation , Automation , Fluorodeoxyglucose F18 , Radioactive Tracers
17.
Appl Radiat Isot ; 113: 89-95, 2016 07.
Article in English | MEDLINE | ID: mdl-27152914

ABSTRACT

For PET radionuclides, the radioactivity of a sample can be conveniently measured by a dose calibrator. These devices depend on a "calibration setting number", but many recommended settings from manuals were interpolated based on standard sources of other radionuclide(s). We conducted HPGe gamma-ray spectroscopy, resulting in a reference for determining settings in two types of vessels containing one of several PET radionuclides. Our results reiterate the notion that in-house, experimental calibrations are recommended for different radionuclides and vessels.


Subject(s)
Positron-Emission Tomography/statistics & numerical data , Positron-Emission Tomography/standards , Radioisotopes/analysis , Radiometry/statistics & numerical data , Radiometry/standards , Bromine Radioisotopes/analysis , Calibration , Copper Radioisotopes/analysis , Humans , Iodine Radioisotopes/analysis , Manganese/analysis , Yttrium Radioisotopes/analysis , Zirconium/analysis
18.
Nuklearmedizin ; 55(4): 172-5, 2016 Aug 05.
Article in German | MEDLINE | ID: mdl-27080701

ABSTRACT

UNLABELLED: Dose calibrators are one of the most important and most frequently used instruments for the determination of activities in nuclear medicine. For guaranteeing a constant quality of the dose calibrators' measurements, constancy checks including the examination of the system linearity have to be performed regularly, usually measured using 99mTc. The 99mTc eluate extracted from a 99Mo/99mTc generator is contaminated with molybdenum. Not accounting for the molybdenum impurity might lead to an exceed of the tolerance limit of 5% deviation to the reference value for this constancy check. The reason for this effect is the contamination of the 99mTc eluate with 99Mo, whose impact depends on both the amount of the impurity and on the total measurement time (high start activities). RESULT: In this work, the influence of the molybdenum impurity on the results of the constancy check of the system linearity was investigated and maximum start activities for certain impurities were determined providing that the deviation to the reference values is below 5%. CONCLUSION: Provided that certain boundary conditions are observed, a correction of the results with respect to the molybdenum contamination is not necessary.


Subject(s)
Drug Contamination , Molybdenum/analysis , Radiometry/instrumentation , Radiometry/standards , Technetium/analysis , Technetium/chemistry , Calibration/standards , Equipment Design , Equipment Failure Analysis , Germany , Molybdenum/chemistry , Nuclear Medicine/standards , Reproducibility of Results , Sensitivity and Specificity
19.
Appl Radiat Isot ; 112: 62-8, 2016 06.
Article in English | MEDLINE | ID: mdl-27015651

ABSTRACT

A general-purpose nuclear medicine dose calibrator was assessed as a potential replacement for a dedicated air-communicating well-type ionization chamber (brachytherapy source strength verification instrument) for (125)I seed source strength verification for radioactive seed localization, where less stringent accuracy tolerances may be acceptable. The accuracy, precision and reproducibility of the dose calibrator were measured and compared to regulatory requirements. The results of this work indicate that a dose calibrator can be used for (125)I seed source strength verification for radioactive seed localization.


Subject(s)
Brachytherapy/methods , Iodine Radioisotopes/administration & dosage , Brachytherapy/statistics & numerical data , Calibration , Humans , Iodine Radioisotopes/therapeutic use , Radiometry/instrumentation , Radiotherapy Dosage , Reproducibility of Results
20.
World J Nucl Med ; 14(3): 165-70, 2015.
Article in English | MEDLINE | ID: mdl-26420986

ABSTRACT

Dose calibrator linearity testing is indispensable for evaluating the capacity of this equipment in measuring radioisotope activities at different magnitudes, a fundamental aspect of the daily routine of a nuclear medicine department, and with an impact on patient exposure. The main aims of this study were to evaluate the feasibility of substituting the radioisotope Fluorine-18 ((18)F) with Technetium-99m ((99m)Tc) in this test, and to indicate it with the lowest operational cost. The test was applied with sources of (99m)Tc (62 GBq) and (18)F (12 GBq), the activities of which were measured at different times, with the equipment preadjusted to measuring sources of (99m)Tc, (18)F, Gallium-67 ((67)Ga), and Iodine-131 ((131)I). Over time, the average deviation between measured and expected activities from (99m)Tc and (18)F were, respectively, 0.56 (±1.79)% and 0.92 (±1.19)%. The average ratios for 99(m)Tc source experimental activity, when measured with the equipment adjusted for measuring (18)F, (67)Ga, and (131)I sources, in real values, were, respectively, 3.42 (±0.06), 1.45 (±0.03), and 1.13 (±0.02), and those for the (18)F source experimental activity, measured through adjustments of (99m)Tc, (67)Ga, and (131)I, were, respectively, 0.295 (±0.004), 0.335 (±0.007), and 0.426 (±0.006). The adjustment of a simple exponential function for describing (99m)Tc and (18)F experimental activities facilitated the calculation of the physical half-lives of the radioisotopes, with a difference of about 1% in relation to the values described in the literature. Linearity test results, when using (99m)Tc, through being compatible with those acquired with (18)F, imply the possibility of using both radioisotopes during linearity testing. Nevertheless, this information, along with the high potential of exposure and the high cost of (18)F, implies that (99m)Tc should preferably be employed for linearity testing in clinics that normally use (18)F, without the risk of prejudicing either the procedure itself or the guarantee of a high-quality nuclear medicine service.

SELECTION OF CITATIONS