Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros








Base de dados
Intervalo de ano de publicação
1.
BMC Ophthalmol ; 24(1): 138, 2024 Mar 27.
Artigo em Inglês | MEDLINE | ID: mdl-38539111

RESUMO

PURPOSE: To assess the level of agreement and evaluate the reliability of measurements between two Scheimpflug imaging modalities, Scansys (MediWorks, China) and Sirius (CSO, Italy), in quantifying the anterior segment parameters in healthy eyes. METHODS: In a cross-sectional study, the right eyes of 38 healthy participants without any ocular or systemic diseases were examined. A range of anterior segment parameters including anterior and posterior flat and steep keratometry, central corneal thickness (CCT), thinnest corneal thickness (TCT), anterior chamber depth (ACD), anterior chamber angle (ACA), corneal volume, anterior chamber volume, and horizontal white to white diameter, derived from the sagittal curvature maps were measured. To evaluate the reliability of the measurements, intraclass correlation coefficient (ICC) and correlation coefficient were measured. Additionally, Bland-Altman plots were employed to examine the agreement in mean (bias line) and 95% limits of agreement between the two devices. RESULTS: The mean age was 31.5 ± 6.9 (range: 19-47) years. The ICC indicated that the majority of anterior segment parameters had an excellent or good level of reliability, surpassing the threshold of 0.9. Nevertheless, CCT and ACA exhibited a moderate level of reliability, with ICC values of 0.794 and 0.728, respectively. The correlation analysis showed a strong correlation for all the variables tested. The Bland-Altman plots revealed that the bias line was near zero and the 95% limits of agreement were narrow for most variables, except for the anterior flat and steep keratometry, which were found to range from - 0.57 to 0.84 D and - 0.68 to 0.87 D, respectively. CONCLUSION: Scansys and Sirius devices can be effectively used interchangeably for the evaluation of most anterior segment parameters; however, for anterior corneal curvatures, CCT and ACA, their alternative use is not recommended.


Assuntos
Córnea , Tomografia , Humanos , Adulto Jovem , Adulto , Reprodutibilidade dos Testes , Estudos Transversais , Estudos Prospectivos , Córnea/diagnóstico por imagem , Topografia da Córnea/métodos , Tomografia de Coerência Óptica
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA