1.
Br J Anaesth ; 86(1): 99-102, 2001 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-11575419

RESUMO

In this study the decontamination procedures of laryngoscopes in Dutch hospitals are described, based on a structured telephone questionnaire. There were substantial differences between decontamination procedures in Dutch hospitals and the standards of the APIC (Association of Professionals in Infection Control and Epidemiology), CDC (Centers of Disease Control) and ASA (American Society of Anesthesiology) were met in full in 19.4% of the hospitals. The standards of manual decontamination, used in 78% of the 139 hospitals, were particularly disappointing; manual cleaning was considered inadequate in 22.9% of these hospitals and manual disinfection did not meet the standards of the APIC, CDC or ASA in any of these hospitals. Decontamination by instrument cleaning machines as a standard procedure was used in 30 (22%) hospitals. In three of these hospitals the blades were subsequently sterilized. We suggest adherence to the infection control guidelines of the CDC, APIC and ASA, until the safety of less conservative infection control practices are demonstrated.


Assuntos
Contaminação de Equipamentos , Laringoscópios , Esterilização/normas , Desinfecção/métodos , Desinfecção/normas , Hospitais Gerais/normas , Humanos , Países Baixos , Guias de Prática Clínica como Assunto , Esterilização/métodos , Inquéritos e Questionários
2.
Anaesthesia ; 54(9): 875-8, 1999 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-10460560

RESUMO

We studied the repeated effect of sterilisation on light intensity in laryngoscopes from Penlon, Riester, Heine (two different blades), Medicon and Upsher. Light intensity was measured by a light meter using two methods. Measurements were performed before the decontamination procedure was carried out and subsequently after each series of 25 procedures until a total of 200 cycles was reached. Using method 1 (and 2), the reduction in light intensity after 200 cycles was 100% (100%; no light emitted), 37% (13%), 75% (69%), 79% (60%), 37% (14%) and 63% (55%) for each blade, respectively.


Assuntos
Tecnologia de Fibra Óptica/instrumentação , Laringoscópios , Laringoscopia , Vapor , Esterilização/métodos , Falha de Equipamento , Humanos , Iluminação