1.
J Assoc Res Otolaryngol ; 20(3): 291-303, 2019 06.
Artigo em Inglês | MEDLINE | ID: mdl-30911952

RESUMO

Cochlear implant (CI) programming is similar for all CI users despite limited understanding of the electrode-neuron interface (ENI). The ENI refers to the ability of each CI electrode to effectively stimulate target auditory neurons and is influenced by electrode position, neural health, cochlear geometry, and bone and tissue growth in the cochlea. Hearing history likely affects these variables, suggesting that the efficacy of each channel of stimulation differs between children who were implanted at young ages and adults who lost hearing and received a CI later in life. This study examined whether ENI quality differed between early-implanted children and late-implanted adults. Auditory detection thresholds and most comfortable levels (MCLs) were obtained with monopolar and focused electrode configurations. Channel-to-channel variability and dynamic range were calculated for both types of stimulation. Electrical field imaging data were also acquired to estimate levels of intracochlear resistance. Children exhibited lower average auditory perception thresholds and MCLs compared with adults, particularly with focused stimulation. However, neither dynamic range nor channel-to-channel threshold variability differed between groups, suggesting that children's range of perceptible current was shifted downward. Children also demonstrated increased intracochlear resistance levels relative to the adult group, possibly reflecting greater ossification or tissue growth after CI surgery. These results illustrate physical and perceptual differences related to the ENI of early-implanted children compared with late-implanted adults. Evidence from this study demonstrates a need for further investigation of the ENI in CI users with varying hearing histories.


Assuntos
Implante Coclear , Implantes Cocleares , Gânglio Espiral da Cóclea/fisiologia , Adolescente , Fatores Etários , Idoso , Idoso de 80 Anos ou mais , Criança , Humanos , Pessoa de Meia-Idade
2.
Int J Aging Res ; 2(2)2019.
Artigo em Inglês | MEDLINE | ID: mdl-34723185

RESUMO

BACKGROUND: The principles of motor learning (PML) emerged from studies of limb motor skills in healthy, young adults. The applicability of these principles to speech motor learning, and to older adults, is uncertain. AIMS: The purpose of this study was to examine one PML, feedback frequency, and its effect on retention and generalization of a novel speech and comparable tracing task. METHODS: Sixty older adults completed a speech motor learning task requiring the production of a novel phrase at speaking rates 2 times and 3 times slower than habitual rate. Participants also completed a limb motor learning task requiring the tracing of a sine wave 2x and 3x slower than habitual rate. Participants were randomly assigned to receive feedback every trial, every 5th trial, or every 10th trial. Mean absolute error was measured to examine immediate generalization, delayed generalization, and 2-day retention. FINDINGS: Results suggested that feedback frequency did not have an effect on the retention and generalization of the speech or manual task, supporting the small but growing literature highlighting the constraints of generalizing the PML to other modalities and populations.

3.
Trends Hear ; 202016 May 17.
Artigo em Inglês | MEDLINE | ID: mdl-27194155

RESUMO

The 10 consonant-nucleus-consonant (CNC) word lists are considered the gold standard in the testing of cochlear implant (CI) users. However, variance in scores across lists could degrade the sensitivity and reliability of them to identify deficits in speech perception. This study examined the relationship between variability in performance among lists and the lexical characteristics of the words. Data are from 28 adult CI users. Each subject was tested on all 10 CNC word lists. Data were analyzed in terms of lexical characteristics, lexical frequency, neighborhood density, bi-, and tri-phonemic probabilities. To determine whether individual performance variability across lists can be reduced, the standard set of 10 phonetically balanced 50-word lists was redistributed into a new set of lists using two sampling strategies: (a) balancing with respect to word lexical frequency or (b) selecting words with equal probability. The mean performance on the CNC lists varied from 53.1% to 62.4% correct. The average difference between the highest and lowest scores within individuals across the lists was 20.9% (from 12% to 28%). Lexical frequency and bi-phonemic probabilities were correlated with word recognition performance. The range of scores was not significantly reduced for all individuals when responses were simulated with 1,000 sets of redistributed lists, using both types of sampling methods. These results indicate that resampling of words does not affect the test-retest reliability and diagnostic value of the CNC word test.


Assuntos
Implante Coclear/instrumentação , Implantes Cocleares , Sinais (Psicologia) , Música , Pessoas com Deficiência Auditiva/reabilitação , Percepção da Altura Sonora , Estimulação Acústica , Adulto , Idoso , Audiometria , Estimulação Elétrica , Feminino , Humanos , Percepção Sonora , Masculino , Pessoas com Deficiência Auditiva/psicologia , Desenho de Prótese , Reconhecimento Psicológico
4.
Trends Hear ; 192015 Feb 04.
Artigo em Inglês | MEDLINE | ID: mdl-25656797

RESUMO

A rapid threshold measurement procedure, based on Bekesy tracking, is proposed and evaluated for use with cochlear implants (CIs). Fifteen postlingually deafened adult CI users participated. Absolute thresholds for 200-ms trains of biphasic pulses were measured using the new tracking procedure and were compared with thresholds obtained with a traditional forced-choice adaptive procedure under both monopolar and quadrupolar stimulation. Virtual spectral sweeps across the electrode array were implemented in the tracking procedure via current steering, which divides the current between two adjacent electrodes and varies the proportion of current directed to each electrode. Overall, no systematic differences were found between threshold estimates with the new channel sweep procedure and estimates using the adaptive forced-choice procedure. Test-retest reliability for the thresholds from the sweep procedure was somewhat poorer than for thresholds from the forced-choice procedure. However, the new method was about 4 times faster for the same number of repetitions. Overall the reliability and speed of the new tracking procedure provides it with the potential to estimate thresholds in a clinical setting. Rapid methods for estimating thresholds could be of particular clinical importance in combination with focused stimulation techniques that result in larger threshold variations between electrodes.


Assuntos
Limiar Auditivo , Implante Coclear/instrumentação , Implantes Cocleares , Surdez/reabilitação , Pessoas com Deficiência Auditiva/reabilitação , Psicoacústica , Estimulação Acústica , Adulto , Idoso , Idoso de 80 Anos ou mais , Surdez/diagnóstico , Surdez/psicologia , Estimulação Elétrica , Estudos de Viabilidade , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Pessoas com Deficiência Auditiva/psicologia , Valor Preditivo dos Testes , Desenho de Prótese , Reprodutibilidade dos Testes , Detecção de Sinal Psicológico , Processamento de Sinais Assistido por Computador , Fatores de Tempo
5.
Otol Neurotol ; 35(1): 136-47, 2014 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-24317220

RESUMO

HYPOTHESIS: A functional vestibular prosthesis can be implanted in human such that electrical stimulation of each semicircular canal produces canal-specific eye movements while preserving vestibular and auditory function. BACKGROUND: A number of vestibular disorders could be treated with prosthetic stimulation of the vestibular end organs. We have previously demonstrated in rhesus monkeys that a vestibular neurostimulator, based on the Nucleus Freedom cochlear implant, can produce canal-specific electrically evoked eye movements while preserving auditory and vestibular function. An investigational device exemption has been obtained from the FDA to study the feasibility of treating uncontrolled Ménière's disease with the device. METHODS: The UW/Nucleus vestibular implant was implanted in the perilymphatic space adjacent to the three semicircular canal ampullae of a human subject with uncontrolled Ménière's disease. Preoperative and postoperative vestibular and auditory function was assessed. Electrically evoked eye movements were measured at 2 time points postoperatively. RESULTS: Implantation of all semicircular canals was technically feasible. Horizontal canal and auditory function were largely, but not totally, lost. Electrode stimulation in 2 of 3 canals resulted in canal-appropriate eye movements. Over time, stimulation thresholds increased. CONCLUSION: Prosthetic implantation of the semicircular canals in humans is technically feasible. Electrical stimulation resulted in canal-specific eye movements, although thresholds increased over time. Preservation of native auditory and vestibular function, previously observed in animals, was not demonstrated in a single subject with advanced Ménière's disease.


Assuntos
Doença de Meniere/cirurgia , Implantação de Prótese , Canais Semicirculares/cirurgia , Potenciais Evocados Miogênicos Vestibulares/fisiologia , Vestíbulo do Labirinto/cirurgia , Estimulação Elétrica , Audição/fisiologia , Humanos , Masculino , Doença de Meniere/fisiopatologia , Pessoa de Meia-Idade , Reflexo Vestíbulo-Ocular/fisiologia , Canais Semicirculares/fisiopatologia , Resultado do Tratamento , Vestíbulo do Labirinto/fisiopatologia
6.
IEEE Trans Biomed Eng ; 60(6): 1685-92, 2013 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-23358943

RESUMO

A vestibular neural prosthesis was designed on the basis of a cochlear implant for treatment of Meniere's disease and other vestibular disorders. Computer control software was developed to generate patterned pulse stimuli for exploring optimal parameters to activate the vestibular nerve. Two rhesus monkeys were implanted with the prototype vestibular prosthesis and they were behaviorally evaluated post implantation surgery. Horizontal and vertical eye movement responses to patterned electrical pulse stimulations were collected on both monkeys. Pulse amplitude modulated (PAM) and pulse rate modulated (PRM) trains were applied to the lateral canal of each implanted animal. Robust slow-phase nystagmus responses following the PAM or PRM modulation pattern were observed in both implanted monkeys in the direction consistent with the activation of the implanted canal. Both PAM and PRM pulse trains can elicit a significant amount of in-phase modulated eye velocity changes and they could potentially be used for efficiently coding head rotational signals in future vestibular neural prostheses.


Assuntos
Implantes Cocleares , Estimulação Elétrica/métodos , Implantes Experimentais , Processamento de Sinais Assistido por Computador/instrumentação , Animais , Eletrodos , Potenciais Evocados/fisiologia , Movimentos Oculares/fisiologia , Macaca mulatta , Desenho de Prótese , Vestíbulo do Labirinto/cirurgia
7.
Hear Res ; 287(1-2): 51-6, 2012 May.
Artigo em Inglês | MEDLINE | ID: mdl-22504025

RESUMO

We measured auditory brainstem responses (ABRs) in eight Rhesus monkeys after implantation of electrodes in the semicircular canals of one ear, using a multi-channel vestibular prosthesis based on cochlear implant technology. In five animals, click-evoked ABR thresholds in the implanted ear were within 10 dB of thresholds in the non-implanted control ear. Threshold differences in the remaining three animals varied from 18 to 69 dB, indicating mild to severe hearing losses. Click- and tone-evoked ABRs measured in a subset of animals before and after implantation revealed a comparable pattern of threshold changes. Thresholds obtained five months or more after implantation--a period in which the prosthesis regularly delivered electrical stimulation to achieve functional activation of the vestibular system--improved in three animals with no or mild initial hearing loss and increased in a fourth with a moderate hearing loss. These results suggest that, although there is a risk of hearing loss with unilateral vestibular implantation to treat balance disorders, the surgery can be performed in a manner that preserves hearing over an extended period of functional stimulation.


Assuntos
Implante Coclear/instrumentação , Implantes Cocleares , Canais Semicirculares/inervação , Vestíbulo do Labirinto/inervação , Estimulação Acústica , Animais , Limiar Auditivo , Implante Coclear/efeitos adversos , Estimulação Elétrica , Eletroencefalografia , Potenciais Evocados Auditivos do Tronco Encefálico , Movimentos Oculares , Perda Auditiva/etiologia , Perda Auditiva/fisiopatologia , Macaca mulatta , Masculino , Desenho de Prótese , Tempo de Reação , Medição de Risco , Fatores de Tempo
8.
Artigo em Inglês | MEDLINE | ID: mdl-23367327

RESUMO

Electrical stimulation of the vestibular end organ with a vestibular prosthesis may provide an effective treatment for vestibular loss if the stimulation remains effective over a significant period of time after implantation of the device. To assess efficacy of electrical stimulation in an animal model, we implanted 3 rhesus monkeys with a vestibular prosthesis based on a cochlear implant. We then recorded vestibular electrically evoked compound action potentials (vECAPs) longitudinally in each of the implanted canals to see how the amplitude of the response changed over time. The results suggest that vECAPs, and therefore electrical activation of vestibular afferent fibers, can remain largely stable over time following implantation.


Assuntos
Potenciais de Ação , Próteses e Implantes , Animais , Estudos Longitudinais , Macaca mulatta , Vestíbulo do Labirinto/cirurgia
9.
Otol Neurotol ; 32(1): 88-97, 2011 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-21192375

RESUMO

OBJECTIVE: We recorded intraoperative and postoperative electrically evoked compound action potentials (ECAPs) in rhesus monkeys implanted with a vestibular neurostimulator. The objectives were to correlate the generation of slow-phase nystagmus or eye twitches induced by electrical stimulation of the implanted semicircular canal with the presence or absence of the vestibular ECAP responses and to assess the effectiveness of ECAP monitoring during surgery to guide surgical insertion of electrode arrays into the canals. DESIGN: Four rhesus monkeys (a total of 7 canals) were implanted with a vestibular neurostimulator modified from the Nucleus Freedom cochlear implant. ECAP recordings were obtained during surgery or at various intervals after surgery using the Neural Response Telemetry feature of the clinical Custom Sound EP software. Eye movements during electrical stimulation of individual canals were recorded with a scleral search coil system in the same animals. RESULTS: Measurable vestibular ECAPs were observed intraoperatively or postoperatively in 3 implanted animals. Robust and sustained ECAPs were obtained in 3 monkeys at the test intervals of 0, 7, or greater than 100 days after implantation surgery. In all 3 animals, stimulation with electrical pulse trains produced measurable eye movements in a direction consistent with the vestibulo-ocular reflex from the implanted semicircular canal. In contrast, electrically evoked eye movements could not be measured in 3 of the 7 implanted canals, none of which produced distinct vestibular ECAPs. In 2 animals, ECAP waveforms were systematically monitored during surgery, and the procedure proved crucial to the success of vestibular implantation. CONCLUSION: Vestibular ECAPs exhibit similar morphology and growth characteristics to cochlear ECAPs from human cochlear implant patients. The ECAP measure is well correlated with the functional activation of eye movements by electrical stimulation after implantation surgery. The intraoperative ECAP recording technique is an efficient tool to guide the placement of electrode array into the semicircular canals.


Assuntos
Potenciais de Ação/fisiologia , Potenciais Evocados/fisiologia , Nervo Vestibular/fisiologia , Animais , Estimulação Elétrica , Eletrodos Implantados , Macaca mulatta , Nistagmo Fisiológico/fisiologia , Reflexo Vestíbulo-Ocular/fisiologia
10.
Artigo em Inglês | MEDLINE | ID: mdl-22255103

RESUMO

Loss of vestibular function results in imbalance, disorientation, and oscillopsia. Several groups have designed and constructed implantable devices to restore vestibular function through electrical stimulation of the vestibular nerve. We have designed a two-part device in which the head motion sensing and signal processing elements are externally mounted to the head, and are coupled through an inductive link to a receiver stimulator that is based on a cochlear implant. The implanted electrode arrays are designed to preserve rotational sensitivity in the implanted ear. We have tested the device in rhesus monkeys by rotating the animals in the plane of the implanted canals, and then using head velocity and acceleration signals to drive electrical stimulation of the vestibular system. Combined electrical and rotational stimulation results in a summation of responses, so that one can control the modulation of eye velocity induced by sinusoidal yaw rotation.


Assuntos
Próteses e Implantes , Nervo Vestibular/fisiopatologia , Animais , Implantes Cocleares , Macaca mulatta
11.
Hear Res ; 270(1-2): 134-42, 2010 Dec 01.
Artigo em Inglês | MEDLINE | ID: mdl-20727397

RESUMO

This study examines patterns of neural activity in response to single biphasic electrical pulses, presented alone or following a forward masking pulse train, delivered by a cochlear implant. Recordings were made along the tonotopic axis of the central nucleus of the inferior colliculus (ICC) in ketamine/xylazine anesthetized guinea pigs. The partial tripolar electrode configuration was used, which provided a systematic way to vary the tonotopic extent of ICC activation between monopolar (broad) and tripolar (narrow) extremes while maintaining the same peak of activation. The forward masking paradigm consisted of a 200 ms masker pulse train (1017 pulses per second) followed 10 ms later by a single-pulse probe stimulus; the current fraction of the probe was set to 0 (monopolar), 1 (tripolar), or 0.5 (hybrid), and the fraction of the masker was fixed at 0.5. Forward masking tuning profiles were derived from the amount of masking current required to just suppress the activity produced by a fixed-level probe. These profiles were sharper for more focused probe configurations, approximating the pattern of neural activity elicited by single (non-masked) pulses. The result helps to bridge the gap between previous findings in animals and recent psychophysical data.


Assuntos
Vias Auditivas/fisiologia , Percepção Auditiva , Implantes Cocleares , Colículos Inferiores/fisiologia , Mascaramento Perceptivo , Estimulação Acústica , Animais , Limiar Auditivo , Estimulação Elétrica , Cobaias , Fatores de Tempo
12.
Hear Res ; 268(1-2): 93-104, 2010 Sep 01.
Artigo em Inglês | MEDLINE | ID: mdl-20580801

RESUMO

The partial tripolar electrode configuration is a relatively novel stimulation strategy that can generate more spatially focused electric fields than the commonly used monopolar configuration. Focused stimulation strategies should improve spectral resolution in cochlear implant users, but may also be more sensitive to local irregularities in the electrode-neuron interface. In this study, we develop a practical computer model of cochlear implant stimulation that can simulate neural activation in a simplified cochlear geometry and we relate the resulting patterns of neural activity to basic psychophysical measures. We examine how two types of local irregularities in the electrode-neuron interface, variations in spiral ganglion nerve density and electrode position within the scala tympani, affect the simulated neural activation patterns and how these patterns change with electrode configuration. The model shows that higher partial tripolar fractions activate more spatially restricted populations of neurons at all current levels and require higher current levels to excite a given number of neurons. We find that threshold levels are more sensitive at high partial tripolar fractions to both types of irregularities, but these effects are not independent. In particular, at close electrode-neuron distances, activation is typically more spatially localized which leads to a greater influence of neural dead regions.


Assuntos
Implantes Cocleares , Nervo Coclear/patologia , Simulação por Computador , Modelos Neurológicos , Rampa do Tímpano/inervação , Gânglio Espiral da Cóclea/patologia , Sobrevivência Celular , Estimulação Elétrica , Humanos , Análise Numérica Assistida por Computador , Desenho de Prótese , Psicoacústica , Limiar Sensorial
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA