Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros








Base de dados
Intervalo de ano de publicação
1.
Radiat Oncol ; 7: 24, 2012 Feb 20.
Artigo em Inglês | MEDLINE | ID: mdl-22348595

RESUMO

PURPOSE: To evaluate gold marker displacement due to needle insertion during HDR-brachytherapy for therapy of prostate cancer. PATIENTS AND METHODS: 18 patients entered into this prospective evaluation. Three gold markers were implanted into the prostate during the first HDR-brachytherapy procedure after the irradiation was administered. Three days after marker implantation all patients had a CT-scan for planning purpose of the percutaneous irradiation. Marker localization was defined on the digitally-reconstructed-radiographs (DRR) for daily (VMAT technique) or weekly (IMRT) set-up error correction. Percutaneous therapy started one week after first HDR-brachytherapy. After the second HDR-brachytherapy, two weeks after first HDR-brachtherapy, a cone-beam CT-scan was done to evaluate marker displacement due to needle insertion. In case of marker displacement, the actual positions of the gold markers were adjusted on the DRR. RESULTS: The value of the gold marker displacement due to the second HDR-brachytherapy was analyzed in all patients and for each gold marker by comparison of the marker positions in the prostate after soft tissue registration of the prostate of the CT-scans prior the first and second HDR-brachytherapy. The maximum deviation was 5 mm, 7 mm and 12 mm for the anterior-posterior, lateral and superior-inferior direction. At least one marker in each patient showed a significant displacement and therefore new marker positions were adjusted on the DRRs for the ongoing percutaneous therapy. CONCLUSIONS: Needle insertion in the prostate due to HDR-brachytherapy can lead to gold marker displacements. Therefore, it is necessary to verify the actual position of markers after the second HDR-brachytherapy. In case of significant deviations, a new DRR with the adjusted marker positions should be generated for precise positioning during the ongoing percutaneous irradiation.


Assuntos
Braquiterapia/instrumentação , Tomografia Computadorizada de Feixe Cônico , Radioisótopos de Ouro , Neoplasias da Próstata/diagnóstico por imagem , Neoplasias da Próstata/radioterapia , Próteses e Implantes , Planejamento da Radioterapia Assistida por Computador , Idoso , Humanos , Masculino , Pessoa de Meia-Idade , Agulhas , Estudos Prospectivos , Intensificação de Imagem Radiográfica
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA