This paper presents the development of a novel high-pressure/high-temperature reactor cell dedicated to the characterization of catalysts using synchrotron x-ray absorption spectroscopy under operando conditions. The design of the vitreous carbon reactor allows its use as a plug-flow reactor, monitoring catalyst samples in a powder form with a continuous gas flow at high-temperature (up to 1000 °C) and under high pressure (up to 1000 bar) conditions, depending on the gas environment. The high-pressure/high-temperature reactor cell incorporates an automated gas distribution system and offers the capability to operate in both transmission and fluorescence detection modes. The operando x-ray absorption spectroscopy results obtained on a bimetallic InCo catalyst during CO2 hydrogenation reaction at 300 °C and 50 bar are presented, replicating the conditions of a conventional microreactor. The complete setup is available for users and permanently installed on the Collaborating Research Groups French Absorption spectroscopy beamline in Material and Environmental (CRG-FAME) sciences and French Absorption spectroscopy beamline in Material and Environmental sciences at ultra-high dilution (FAME-UHD) beamlines (BM30 and BM16) at the European Synchrotron Radiation Facility in Grenoble, France.