RESUMO
Numerous biomaterials have been developed for application in blood-contacting medical devices to prevent thrombosis; however, few materials have been applied to full-scale devices and evaluated for hemocompatibility under clinical blood flow conditions. We applied a dual-action slippery liquid-infused (LI) nitric oxide (NO)-releasing material modification (LINO) to full-scale blood circulation tubing for extracorporeal lung support and evaluated the tubing ex vivo using swine whole blood circulated for 6 h at a clinically relevant flow. LINO tubing was compared to unmodified tubing (CTRL) and isolated LI and NO-releasing modifications (n = 9/group). The primary objective was to evaluate safety and blood compatibility of this approach, prior to progression to in vivo testing of efficacy in animal models. The secondary objective was to evaluate coagulation outcomes relevant to hemocompatibility. No untoward effects of the coating, such as elevated methemoglobin fraction, were observed. Additionally, LINO delayed platelet loss until 6 h versus the reduction in platelet count in CTRL at 3 h. At 6 h, LINO significantly reduced the concentration of platelets in an activated P-selectin expressing state versus CTRL (32 ± 1% decrease, p = .02). Blood clot deposition was significantly reduced on LINO blood pumps (p = .007) and numerically reduced on tubing versus CTRL. Following blood exposure, LINO tubing continued to produce a measurable NO-flux (0.20 ± 0.06 × 10-10 mol cm-2 min-1 ). LINO is a potential solution to reduce circuit-related bleeding and clotting during extracorporeal organ support, pending future extended testing in vivo using full-scale extracorporeal lung support devices.