Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros








Base de dados
Intervalo de ano de publicação
1.
Zhongguo Yi Xue Ke Xue Yuan Xue Bao ; 31(2): 151-4, 2009 Apr.
Artigo em Chinês | MEDLINE | ID: mdl-19507591

RESUMO

OBJECTIVE: To investigate the value of in vivo proton magnetic resonance spectroscopy (1H MRS) in the assessment of hepatocellular carcinoma (HCC) and cholangiocarcinoma. METHODS: 1H MRS was performed in normal volunteers and in patients with pathologically confirmed HCC and cholangiocarcinomas using a whole-body 1.5-T scanner. The choline-to-lipid ratios were measured by dividing the peak area of choline at 3.2 ppm and lipid at 1.3 ppm. RESULTS: The ratio of choline-to-lipid for normal liver, cholangiocarcinomas, and HCC were 0.07 +/- 0.04, 0.11 +/- 0.06, and 0.52 +/- 0.15, respectively. The ratio of choline-to-lipid was significantly higher in HCC compared than those in cholangiocarcinomas or normal livers (P < 0.05). However, it was not significantly different between cholangiocarcinomas and normal livers (P > 0.05). CONCLUSION: In vivo 1H MRS can reflect the pathological changes of HCC and cholangiocarcinomas at metabolic level and thus is useful in the diagnosis of these two cancers.


Assuntos
Neoplasias dos Ductos Biliares/diagnóstico , Ductos Biliares Intra-Hepáticos , Carcinoma Hepatocelular/diagnóstico , Colangiocarcinoma/diagnóstico , Neoplasias Hepáticas/diagnóstico , Espectroscopia de Ressonância Magnética/métodos , Adulto , Idoso , Neoplasias dos Ductos Biliares/metabolismo , Ductos Biliares Intra-Hepáticos/metabolismo , Ductos Biliares Intra-Hepáticos/patologia , Carcinoma Hepatocelular/metabolismo , Estudos de Casos e Controles , Colangiocarcinoma/metabolismo , Feminino , Humanos , Neoplasias Hepáticas/metabolismo , Imageamento por Ressonância Magnética/métodos , Masculino , Pessoa de Meia-Idade , Prótons , Estudos Retrospectivos
2.
Chin Med J (Engl) ; 122(8): 911-6, 2009 Apr 20.
Artigo em Inglês | MEDLINE | ID: mdl-19493413

RESUMO

BACKGROUND: This study was to investigate the safety of two types of commercially available lumbar artificial discs (CHARITE and PRODISC-L) during a magnetic resonance imaging (MRI) procedure in a 1.5-Tesla MR system, and to evaluate the size of metal artifacts on the MR image for different sequences. METHODS: A 1.5-Tesla clinical MR imaging system was used. The degree of deflection of the endplates of two artificial discs was evaluated by an angle-measurement instrument at the portals of the MRI scanner. The heating effect of the radio frequency (RF) magnetic field was evaluated by using "worst-case" imaging sequences on a human cadaver implanted with an artificial lumbar disc at the L5/S1 intervertebral disc location. The temperatures of the tissue adjacent to the implant, and of the L4/L5 intervertebral disc (used as a control) were measured, respectively, using a digital probe thermometer before and after the MRI scan sequence. A rectangular water phantom was designed to evaluate the metal artifacts of these two artificial discs under different MR imaging sequences. RESULTS: The maximal deflection angle of the endplate of the implants under a static MR field was 7.5 and 6.0 degrees, for the CHARITE and PRODISC-L, respectively. The difference between temperature rise of tissue adjacent to the two types of artificial discs and the temperature rise of the L4/L5 control location was 0.4 and 0.6 degrees C, respectively. The size of metal artifacts on images of TSE (T1/T2 -weighted), STIR and Turbo Dark Fluid sequences were relatively less than those of TSE fat saturation, Flash and SE (T1-weighted) sequences. CONCLUSIONS: The CHARITE and the PRODISC-L artificial disc do not present an additional hazard or risk to a patient undergoing an MRI procedure using a scanner operating with a static magnetic field of 1.5 T or lower. Image artifacts from the implants may present problems if the anatomical region of interest is in or near the area where these implants are located (e.g., vertebral canal at affected segment).


Assuntos
Deslocamento do Disco Intervertebral/cirurgia , Disco Intervertebral/cirurgia , Vértebras Lombares/cirurgia , Imageamento por Ressonância Magnética/métodos , Metais , Próteses e Implantes , Segurança de Equipamentos , Estudos de Avaliação como Assunto , Humanos , Disco Intervertebral/patologia , Deslocamento do Disco Intervertebral/patologia , Vértebras Lombares/patologia
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA