Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros








Base de dados
Intervalo de ano de publicação
1.
IEEE Trans Biomed Eng ; 42(7): 731-5, 1995 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-7622157

RESUMO

A method for line interference reduction to be used in signal-averaged electrocardiography (SAECG) systems is proposed and its performance is analyzed. This new method is an adaptation of a previously reported technique for removal of line interference from conventional electrocardiograms. It involves the recording of a line reference signal simultaneous with the lead signals, so that a shifted and scaled version of it can be used to subtract line interference from the leads. It is shown that this line interference subtraction method can reduce line interference effectively and without introducing any additional noise into the ECG signal. It is also shown that Late Potential diagnostic decisions are not altered when this filter is applied. It is recommended that this technique be used in SAECG when line interference is unavoidable.


Assuntos
Eletrocardiografia/instrumentação , Processamento de Sinais Assistido por Computador , Filtração , Humanos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA