Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros








Base de dados
Assunto principal
Intervalo de ano de publicação
1.
Photochem Photobiol Sci ; 8(4): 516-27, 2009 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-19337666

RESUMO

A variety of instruments have been developed over the past 50 years to measure spectral radiance in absolute units at UV and visible wavelengths with high spectral resolution. While there is considerable experience in the measurement of spectral irradiance, less emphasis has been given to the reliable measurement of spectral radiance from ground observations. We discuss the methodology and calibration procedures for five instruments capable of making such measurements. Four of these instruments are based on double monochromators that scan each wavelength in turn, and one is based on a single monochromator with a charged coupled device (CCD) allowing the recording of all wavelengths simultaneously. The measured spectral radiance deviates between 3% and about 35% depending on the instruments. The results are compared with radiative transfer calculations when the aerosol characteristics of the atmosphere are known.


Assuntos
Meteorologia/métodos , Atmosfera , Calibragem/normas , Alemanha , Doses de Radiação , Monitoramento de Radiação/instrumentação , Monitoramento de Radiação/métodos , Espectrofotometria/instrumentação , Espectrofotometria/métodos , Espectrofotometria Ultravioleta/instrumentação , Espectrofotometria Ultravioleta/métodos , Raios Ultravioleta
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA