Viruses ; 16(4)2024 04 13.
Artigo em Inglês | MEDLINE | ID: mdl-38675941

RESUMO

The BIOFIRE SPOTFIRE Respiratory (R) Panel is a novel, in vitro diagnostic PCR assay with 15 pathogen targets. The runtime is about 15 min which is the shortest among similar panels in the market. We evaluated the performance of the SPOTFIRE R Panel with 151 specimens, including 133 collected from the upper respiratory tract (URT), 13 from the lower respiratory tract (LRT) and 5 external quality assessment program (EQAP) samples. The respiratory specimens were enrolled throughout the first two post-COVID-19 influenza seasons in Hong Kong (March to December 2023). For URT specimens, full concordance was observed between the SPOTFIRE R Panel and the standard-of-care FilmArray Respiratory 2.1 plus Panel (RP2.1plus) for 109 specimens (109/133, 81.95%). After discrepant analysis, the SPOTFIRE R Panel identified more pathogens than the RP2.1plus in 15 specimens and vice versa in 3 specimens. The per-target negative and positive percentage agreement (NPA and PPA) were 92.86-100% except the PPA of adenovirus (88.24%). For LRT and EQAP samples, all results were fully concordant. To conclude, the performance of the SPOTFIRE R Panel was comparable to the RP2.1plus.


Assuntos
COVID-19 , Infecções Respiratórias , SARS-CoV-2 , Humanos , COVID-19/diagnóstico , COVID-19/virologia , Infecções Respiratórias/virologia , Infecções Respiratórias/diagnóstico , Hong Kong , SARS-CoV-2/genética , SARS-CoV-2/isolamento & purificação , Sensibilidade e Especificidade , Técnicas de Diagnóstico Molecular/métodos , Teste de Ácido Nucleico para COVID-19/métodos
2.
Diagnostics (Basel) ; 14(7)2024 Mar 24.
Artigo em Inglês | MEDLINE | ID: mdl-38611596

RESUMO

The performance of the Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assays were evaluated using 167 specimens, including 158 human respiratory specimens and 9 external quality assessment program (EQAP) samples. For respiratory specimens, CoV-2/Flu/RSV plus exhibited perfect agreement with the standard-of-care (SOC) methods (Cohen's κ: 1, 100% agreement). The overall positive and negative percent agreement (PPA and NPA) were 100%, with 95% confidence intervals of 96.50 to 100% and 85.70 to 100%, respectively. On the other hand, Resp-4-Plex revealed an almost perfect agreement with the SOC methods (Cohen's κ: 0.92, 97.71% agreement). The overall PPA and NPA were 100% (95.76 to 100%) and 88.46% (70.20 to 96.82%), respectively. For EQAP samples, the results of CoV-2/Flu/RSV plus (9/9) and Resp-4-Plex (4/4) were concordant with the expected results. The experimental limit of detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) was the lowest (25 copies/mL for both methods), and that of the respiratory syncytial virus was the highest (400 copies/mL for CoV-2/Flu/RSV plus and 100 copies/mL for Resp-4-Plex). Threshold cycle (Ct) value correlation showed a large positive linear association between CoV-2/Flu/RSV plus and Resp-4-Plex, with R-squared values of 0.92-0.97, and on average, the Ct values of CoV-2/Flu/RSV plus were higher than that of Resp-4-Plex by 1.86-2.78, except for Flu A1 target (-0.66). To conclude, the performance of both assay was comparable to the SOC methods for both upper and lower respiratory specimens. Implementation of these rapid assay may reinforce the diagnostic capacity for the post-pandemic co-circulation of SARS-CoV-2 and other respiratory viruses.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA