1.
Nat Biotechnol ; 34(3): 320-7, 2016 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-26854476

RESUMO

High-fidelity intracranial electrode arrays for recording and stimulating brain activity have facilitated major advances in the treatment of neurological conditions over the past decade. Traditional arrays require direct implantation into the brain via open craniotomy, which can lead to inflammatory tissue responses, necessitating development of minimally invasive approaches that avoid brain trauma. Here we demonstrate the feasibility of chronically recording brain activity from within a vein using a passive stent-electrode recording array (stentrode). We achieved implantation into a superficial cortical vein overlying the motor cortex via catheter angiography and demonstrate neural recordings in freely moving sheep for up to 190 d. Spectral content and bandwidth of vascular electrocorticography were comparable to those of recordings from epidural surface arrays. Venous internal lumen patency was maintained for the duration of implantation. Stentrodes may have wide ranging applications as a neural interface for treatment of a range of neurological conditions.


Assuntos
Procedimentos Endovasculares , Córtex Motor/fisiologia , Neurônios/fisiologia , Stents , Animais , Catéteres , Angiografia Cerebral/métodos , Eletrodos , Humanos , Ovinos
SELEÇÃO DE REFERÊNCIAS