Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros








Base de dados
Intervalo de ano de publicação
1.
Anal Sci ; 40(6): 1071-1080, 2024 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-38483788

RESUMO

In this paper, we optimized a method for fast and accurate determination of five impurity elements (As, Sb, Bi, Se, and Ge) in graphite samples to overcome the shortcomings of existing methods, such as complicated equipment, cumbersome process, multiple-time preparation, separate determination, and large error in results. Graphite samples were digested with HNO3-H2SO4-HClO4-HF in a high-temperature and high-pressure microwave digestion apparatus, and the elements were extracted and determined separately by AFS (atomic fluorescence spectrometry). There is no element loss during the processing and analysis of this method. The spike recoveries (As: 90.30%-102.3%, Sb: 90.73%-110.0%, Bi: 90.00%-99.67%, Se: 93.33%-110.0%, Ge: 92.26%-104.2%) and precision (RSD%; As: 1.34%-8.96%, Sb: 2.67%-7.10%, Bi: 1.83%-4.58%, Se: 0.36%-3.25%, Ge: 4.41%-8.65%) meet the requirements of the corresponding quality specifications. The method has some advantages (such as no elemental loss, fast testing, strong element targeting, and accurate results), and thus can achieve batch determination of graphite samples. The optimized method for graphite sample and final solution preparations can be used for diverse spectrometric technologies, and that for spectrometer conditions have reference value for HG-AFS instruments.