Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros








Base de dados
Intervalo de ano de publicação
1.
J Appl Clin Med Phys ; 6(3): 133-42, 2005.
Artigo em Inglês | MEDLINE | ID: mdl-16143798

RESUMO

The BANG polymer gel dosimeter was used to evaluate 3D absorbed dose distributions in tissue delivered with Gamma Knife stereotactic radiosurgery systems. We compared dose distributions calculated with Leksell GammaPlan (LGP) treatment-planning software with dose distributions measured with the polymer gel dosimeter for single-shot irradiations. Head-sized spherical glass vessels filled with the polymer gel were irradiated with Gamma Knife. The phantoms were scanned with a 1.0T MRI scanner. The Hahn spin-echo sequence with two echoes was used for the MRI scans. Calibration relations between the spin-spin relaxation rate and the absorbed dose were obtained by using small cylindrical vials, which were filled with the polymer gel from the same batch as for the spherical phantom. We made voxel-by-voxel comparisons of measured and calculated dose distributions for 31 x 31 x 31 dose matrix elements. With the 3D dose data we calculated the tumor control probability (TCP) and normal tissue complication probability (NTCP) for a simple model. For the maximum dose of 100 Gy, the mean and one standard deviation of differences between the measured and the calculated doses were the following: -0.38+/-4.63 Gy, 1.49+/-2.77 Gy, and -1.03+/-4.18 Gy for 8-mm, 14-mm, and 18-mm collimators, respectively. Tumor control probability values for measurements were smaller than the calculations by 0% to 7%, whereas NTCP values were larger by 7% to 24% for four of six experiments.


Assuntos
Análise de Falha de Equipamento/instrumentação , Polímeros/efeitos da radiação , Garantia da Qualidade dos Cuidados de Saúde/métodos , Radiometria/instrumentação , Radiometria/métodos , Radiocirurgia/instrumentação , Análise de Falha de Equipamento/métodos , Doses de Radiação , Reprodutibilidade dos Testes , Espalhamento de Radiação , Sensibilidade e Especificidade
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA