Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 3 de 3
Filtrar
Mais filtros








Base de dados
Intervalo de ano de publicação
1.
Front Med Biol Eng ; 1(3): 217-27, 1989.
Artigo em Inglês | MEDLINE | ID: mdl-2486762

RESUMO

Since a whole body PET-scanner POSITOLOGICA II was installed at the National Institute of Radiological Sciences (NIRS) in 1982, improvements of the computer system and developments of new software have been continuously made to correspond to clinical requirements. As the result, POSITOLOGICA II possesses one of the most powerful systems and sophisticated software in Japan. This report describes the present status of the computer system and software of POSITOLOGICA II. It also discusses problems which still exist and proposes a future computer system for the PET study.


Assuntos
Sistemas Computacionais , Software , Tomografia Computadorizada de Emissão , Encéfalo/anatomia & histologia , Infarto Cerebral/diagnóstico , Humanos , Processamento de Imagem Assistida por Computador , Computação Matemática , Monitorização Fisiológica , Valores de Referência
2.
Radioisotopes ; 36(5): 221-6, 1987 May.
Artigo em Japonês | MEDLINE | ID: mdl-3497411

RESUMO

This paper describes evaluation and correction of count rate characteristics of POSITOLOGICA II, a multi-slice whole body positron emission tomography system. The present study was performed using three phantoms; a 5 cm inner diameter, water-filled lucite cylinder, a 20 cm inner diameter, water-filled lucite cylinder and a chest phantom. After injection of high activity (about 1.85 GBq (50 mCi] of 13N ammonia into each phantom, rates of true coincidence, random coincidence and single photon detections were measured during decay of the isotope through more than two orders of magnitude of activity. At very high levels of activity, count rate characteristics of the system were saturated and limited to 660 kcps of total coincidence rate, which was the sum of rates in on-time and off-time windows, by the FIFO (first-in first-out) output frequency. Below those levels of activity the relationship between count loss and true coincidence rate was not unique but depended on the phantom configurations, suggesting that count loss correction using the above relationship was inadequate for quantitative study. However, the relationship between count loss and single rate was almost independent of the phantom configurations. Thus in conclusion count loss could be corrected using single rate for POSITOLOGICA II. A practical method of count loss correction was also proposed.


Assuntos
Tomografia Computadorizada de Emissão/instrumentação , Modelos Estruturais
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA