1.
Rev Sci Instrum ; 93(12): 123502, 2022 Dec 01.
Artigo em Inglês | MEDLINE | ID: mdl-36586943

RESUMO

A highly adaptable and robust terahertz (THz) energy meter is designed and implemented to detect energetic THz pulses from high-intensity (>1018 W/cm2) laser-plasma interactions on the OMEGA EP. THz radiation from the laser driven target is detected by a shielded pyrometer. A second identical pyrometer is used for background subtraction. The detector can be configured to detect THz pulses in the 1 mm to 30 µm (0.3- to 10-THz) range and pulse energies from joules to microjoules via changes in filtration, aperture size, and position. Additional polarization selective filtration can also be used to determine the THz pulse polarization. The design incorporates significant radiation and electromagnetic pulse shielding to survive and operate within the OMEGA EP radiation environment. We describe the design, operational principle, calibration, and testing of the THz energy meter. The pyrometers were calibrated using a benchtop laser and show linear sensitivity to up to 1000 nJ of absorbed energy. The initial results from four OMEGA EP THz experiments detected up to ∼15µJ at the detector, which can correspond to hundreds of mJ depending on THz emission and reflection models.