1.
Appl Radiat Isot
; 156: 108935, 2020 Feb.
Artigo
em Inglês
| MEDLINE
| ID: mdl-31675614
RESUMO
Technetium-99m was standardised by the 4π(LS)ce-γ coincidence extrapolation method. Sensitivity of the 4π(LS) channel to two types of radiation, namely conversion electrons and γ-rays, resulted in incorrect activity values being obtained when this was not adequately accounted for. Measurements were more robust when the LS detection efficiency was optimised, and when a γ-window setting was used that monitored the combined LS efficiency for conversion electrons and γ-rays. The primary standard was internationally compared through participation in the BIPM.RI(II)-K4.Tc99m key comparison.