1.
J Bone Joint Surg Br ; 93(12): 1610-6, 2011 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-22161922

RESUMO

Since the Oxford knee was first used unicompartmentally in 1982, a small number of bearings have fractured. Of 14 retrieved bearings, we examined ten samples with known durations in situ (four Phase 1, four Phase 2 and two Phase 3). Evidence of impingement and associated abnormally high wear (> 0.05 mm per year) as well as oxidation was observed in all bearings. In four samples the fracture was associated with the posterior radio-opaque wire. Fracture surfaces indicated fatigue failure, and scanning electron microscopy suggested that the crack initiated in the thinnest region. The estimated incidence of fracture was 3.20% for Phase 1, 0.74% for Phase 2, 0.35% for Phase 3, and 0% for Phase 3 without the posterior marker wire. The important aetiological factors for bearing fracture are impingement leading to high wear, oxidation, and the posterior marker wire. With improved surgical technique, impingement and high wear should be prevented and modern polyethylene may reduce the oxidation risk. A posterior marker wire is no longer used in the polyethylene meniscus. Therefore, the rate of fracture, which is now very low, should be reduced to a negligible level.


Assuntos
Artroplastia do Joelho/instrumentação , Análise de Falha de Equipamento/estatística & dados numéricos , Prótese do Joelho , Falha de Prótese , Idoso , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Desenho de Prótese , Falha de Tratamento
2.
J Bone Joint Surg Br ; 92(3): 367-73, 2010 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-20190307

RESUMO

The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years (sd 4.1), with 20 having been implanted for over ten years. The macroscopic pattern of polyethylene wear and the linear penetration were recorded for each bearing. The mean rate of linear penetration was 0.07 mm/year. The patterns of wear fell into three categories, each with a different rate of linear penetration; 1) no abnormal macroscopic wear and a normal articular surface, n = 16 (linear penetration rate = 0.01 mm/year); 2) abnormal macroscopic wear and normal articular surfaces with extra-articular impingement, n = 16 (linear penetration rate = 0.05 mm/year); 3) abnormal macroscopic wear and abnormal articular surfaces with intra-articular impingement +/- signs of non-congruous articulation, n = 15 (linear penetration rate = 0.12 mm/year). The differences in linear penetration rate were statistically significant (p < 0.001). These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly.


Assuntos
Artroplastia do Joelho/métodos , Prótese do Joelho , Falha de Prótese , Idoso , Idoso de 80 Anos ou mais , Análise de Falha de Equipamento/métodos , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Osteoartrite do Joelho/cirurgia , Polietileno , Desenho de Prótese , Propriedades de Superfície , Fatores de Tempo
3.
Knee ; 14(4): 275-9, 2007 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-17524650

RESUMO

The advantages of Unicompartmental Knee Replacement (UKR) over Total Knee Replacement (TKR) includes the preservation of soft tissue as well as bone stock, and better function with improved range of motion and more natural gait. It is therefore believed that the revision of failed UKR to TKR is technically easier than revision of failed TKR. In our study we tested this hypothesis by assessing the reconstruction requirements and early clinical and radiological outcome following the revision of UKR to TKR. During a 15-year period 1060 primary Oxford medial UKR procedures were performed at three centres, 36 of which underwent revision to TKR due to aseptic failure. The mean operating time for revision surgery was 113 min. Among the revision prostheses used, 28 were standard TKRs, six were constrained, and two were semi-constrained. Thirty had no intramedullary stems whereas six had intramedullary stems. In 30 cases reconstruction for bone loss was not required whereas metal augmentation was used in two knees, contained peg defects in the femur were filled with cement in two knees and contained keel defects in the tibia were grafted using the bone from revision cuts in two knees. After a mean follow-up of 2 years, the mean 'total knee score' was 86.3 and the mean functional score was 78.5. These findings suggest that the complexity of operation and complications encountered during Oxford medial UKR revision and the clinical outcome compare favorably with those of TKR revision.


Assuntos
Artroplastia do Joelho/métodos , Prótese do Joelho , Avaliação de Resultados em Cuidados de Saúde , Idoso , Idoso de 80 Anos ou mais , Cimentos Ósseos/uso terapêutico , Feminino , Seguimentos , Humanos , Masculino , Pessoa de Meia-Idade , Desenho de Prótese , Falha de Prótese , Reoperação
4.
J Bone Joint Surg Br ; 87(11): 1488-92, 2005 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-16260664

RESUMO

We present a comparison of the results of the Oxford unicompartmental knee arthroplasty in patients younger and older than 60 years of age. The ten-year all-cause survival of the < 60 years of age group (52) was 91% (95% confidence interval (CI) 12), while in the > or = 60 years of age group (512), the figure was 96% (95% CI 3). For the younger group, the mean Hospital for Special Surgery score at ten-year follow-up (n = 21) was 94 of 100, compared with a mean of 86 of 100 for the older group (n = 135). The results show that the Oxford unicompartmental arthroplasty can achieve ten-year results that are comparable to total knee arthroplasty in patients < 60 years of age. We conclude that for patients aged over 50, age should not be considered a contraindication for this procedure.


Assuntos
Artroplastia do Joelho/métodos , Osteoartrite do Joelho/cirurgia , Adulto , Fatores Etários , Idoso , Idoso de 80 Anos ou mais , Feminino , Seguimentos , Humanos , Masculino , Pessoa de Meia-Idade , Reoperação , Índice de Gravidade de Doença , Análise de Sobrevida , Resultado do Tratamento
5.
Clin Orthop Relat Res ; (435): 171-80, 2005 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-15930935

RESUMO

UNLABELLED: The medial Oxford unicompartmental knee arthroplasty uses a fully congruent mobile bearing, varying in thickness from 3.5-11.5 mm. Impressive clinical and survival results have been achieved by the designing surgeon, but can they be reproduced by an independent center and does the use of bearings less than 6 mm thick reduce the implant's survival? Between November 1983 and May 2000, 439 medial Oxford prostheses were implanted by three surgeons in Skovde, Sweden. One hundred fourteen knees (89 patients) were reviewed clinically at a minimum of 10 years postoperatively, using the Hospital for Special Surgery knee score. Ninety-one percent of the knees had good or excellent results, with no poor results, and 82% were reported as pain free. The 15-year survival rate for the entire cohort was 93%, and for the 432 knees (344 patients) that met the current indications for using the device, the survival rate was 94%. The 10-year survival rate for bearings less than 6 mm thick was 95%, compared with 94% with bearings greater than 6 mm. The results show that an independent center can achieve excellent long-term clinical and survival results in patients implanted with the Oxford prosthesis. Survival of the implant was not reduced by the use of thin polyethylene bearings. LEVEL OF EVIDENCE: Therapeutic study, Level IV (case series--no, or historical, control group). See the Guidelines for Authors for a complete description of levels of evidence.


Assuntos
Artroplastia do Joelho/métodos , Prótese do Joelho , Idoso , Idoso de 80 Anos ou mais , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Desenho de Prótese , Falha de Prótese , Reoperação , Taxa de Sobrevida , Fatores de Tempo , Resultado do Tratamento
6.
J Bone Joint Surg Br ; 83(7): 1034-6, 2001 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-11603518

RESUMO

Satisfactory selection criteria are essential for the successful outcome of unicompartmental knee arthroplasty (UCA). We report the frequency of revision of the Oxford medial unicompartmental arthroplasty in knees previously treated for anteromedial osteoarthritis by high tibial osteotomy (HTO). The combined results from three sources were used to allow statistical analysis of this uncommon subgroup. In the combined series of 631 knees (507 patients) which had medial unicompartmental replacement, 613 were primary procedures and 18 were for a failed HTO. The mean follow-up times of the two groups were similar (5.8 years and 5.4 years, respectively). At review, 19 (3.1%) of the primary procedures and five (27.8%) of those undertaken for a failed HTO had been revised to total knee replacement. Survival analysis revealed the ten-year cumulative survivals to be 96% and 66%, respectively. The log-rank comparison of these survivals revealed a highly significant difference (p < 0.0001). We recommend that the Oxford UCA should not be used in knees which have previously undergone an HTO.


Assuntos
Artroplastia do Joelho/métodos , Osteotomia/métodos , Tíbia/cirurgia , Adulto , Idoso , Idoso de 80 Anos ou mais , Feminino , Seguimentos , Humanos , Masculino , Pessoa de Meia-Idade , Reoperação , Falha de Tratamento
7.
J Bone Joint Surg Br ; 83(2): 191-4, 2001 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-11284563

RESUMO

We describe the outcome of a series of 124 Oxford meniscal-bearing unicompartmental arthroplasties carried out for osteoarthritis of the medial compartment. They had been undertaken more than ten years ago in a non-teaching hospital in Sweden by three surgeons. All the knees had an intact anterior cruciate ligament, a correctable varus deformity and full-thickness cartilage in the lateral compartment. Thirty-seven patients had died; the mean time since operation for the remainder was 12.5 years (10.1 to 15.6). Using the endpoint of revision for any cause, the outcome for every knee was established. Six had been revised (4.8%). At ten years there were 94 knees still at risk and the cumulative survival rate was 95.0% (95% confidence interval 90.8 to 99.3). This figure is similar to that reported by the designers of the prosthesis and to the best published results for independent series of total knee replacement. If patients are selected appropriately, this implant is a reliable treatment for anteromedial osteoarthritis of the knee.


Assuntos
Artroplastia do Joelho/instrumentação , Prótese do Joelho , Idoso , Feminino , Humanos , Masculino , Pessoa de Meia-Idade
SELEÇÃO DE REFERÊNCIAS