1.
J Chromatogr A ; 1497: 64-69, 2017 May 12.
Artigo em Inglês | MEDLINE | ID: mdl-28385265

RESUMO

The Agilent Chip Cube Interface is a microfluidic chip-based technology originally designed for nanospray molecular mass spectrometry in which the sample enrichment, nano-column, tubing, connectors and spray tip were integrated into a single biocompatible chip. Here we describe the hyphenation of the Chip Cube Interface to ICP-MS via modification of the standard HPLC chip design and a new total consumption nebuliser suitable for flow rates as low as 300nLmin-1. The potential of the instrument to eliminate common nanoLC - ICP-MS shortcomings such as leaks, blockages and band-broadening was demonstrated via analysis of cyanocobalamin in equine plasma. The method was linear over three orders of magnitude with an r2 of 0.9999, the peak area repeatability was 1.9% (n=7), and the detection limit was 14ngmL-1. This novel configuration of the Chip Cube Interface coupled to ICP-MS is a suitable platform for the analysis of biomolecules associated with trace metals and speciation applications.


Assuntos
Cromatografia Líquida de Alta Pressão/métodos , Espectrometria de Massas/métodos , Microfluídica/métodos , Vitamina B 12/sangue , Animais , Cavalos/sangue , Limite de Detecção