1.
Can J Ophthalmol ; 54(1): 136-138, 2019 02.
Artigo em Inglês | MEDLINE | ID: mdl-30851768

RESUMO

OBJECTIVE: To compare relative efficiency and chatter of high aspiration and vacuum settings. DESIGN: In vitro laboratory study. METHODS: The John A. Moran Eye Center Laboratories, University of Utah, Salt Lake City, Utah, was the study setting. Porcine nuclei were fixed in formalin and cut into 2.0 mm cubes. Lens cubes were phacoemulsified with Balanced tips at 50 and 60 mL/min aspiration with 500, 600, and 700 mm Hg vacuum with monitored forced infusion. Experiments were conducted at constant torsional power, longitudinal power, and intraocular pressure. RESULTS: No significant change was observed in average chatter across each tested setting. Increasing aspiration rate did not increase efficiency. Increasing vacuum up to 600 mm Hg from 500 mm Hg did not change efficiency. However, increasing vacuum to 700 mm Hg decreased efficiency (p = 0.008 for 500 mm Hg vs 700 mm Hg and p = 0.05 for 600 mm Hg vs 700 mm Hg). Increasing aspiration and increasing vacuum did not significantly improve chatter. CONCLUSIONS: Increasing aspiration above 50 mL/min did not improve phacoemulsification efficiency. Increasing vacuum settings to 700 mm Hg decreases efficiency. Chatter did not significantly change with increasing aspiration and vacuum settings.


Assuntos
Ondas de Choque de Alta Energia/uso terapêutico , Cristalino/cirurgia , Facoemulsificação/métodos , Animais , Modelos Animais de Doenças , Pressão , Sucção , Suínos , Vácuo
2.
J Cataract Refract Surg ; 43(9): 1202-1206, 2017 09.
Artigo em Inglês | MEDLINE | ID: mdl-28991618

RESUMO

PURPOSE: To determine the optimum bottle height, vacuum, aspiration rate, and power settings in the peristaltic mode of the Whitestar Signature Pro machine with Ellips FX tip action (transversal). SETTING: John A. Moran Eye Center Laboratories, University of Utah, Salt Lake City, Utah, USA. DESIGN: Experimental study. METHODS: Porcine lens nuclei were hardened with formalin and cut into 2.0 mm cubes. Lens cubes were emulsified using transversal and fragment removal time (efficiency), and fragment bounces off the tip (chatter) were measured to determine optimum aspiration rate, bottle height, vacuum, and power settings in the peristaltic mode. RESULTS: Efficiency increased in a linear fashion with increasing bottle height and vacuum. The most efficient aspiration rate was 50 mL/min, with 60 mL/min statistically similar. Increasing power increased efficiency up to 90% with increased chatter at 100%. CONCLUSION: The most efficient values for the settings tested were bottle height at 100 cm, vacuum at 600 mm Hg, aspiration rate of 50 or 60 mL/min, and power at 90%.


Assuntos
Facoemulsificação , Animais , Biometria , Extração de Catarata , Núcleo do Cristalino , Cristalino , Facoemulsificação/instrumentação , Facoemulsificação/métodos , Suínos , Ultrassonografia , Vácuo