1.
Anticancer Res ; 18(3B): 2069-76, 1998.
Artigo em Inglês | MEDLINE | ID: mdl-9677469

RESUMO

BACKGROUND: Reproducibility of dose rate at point A is the main characteristic of Manchester system for dose prescription for cancer cervix brachytherapy. Here, we investigated dose rate variations at point A in the application of Selectron-MDR and microSelectron-HDR. MATERIALS AND METHODS: All applications were compared for their physical aspects (loading and geometry). Two points A(rev) and A(or) (revised and original definitions) were checked for dose rate variations. Here we proposed a tool "dose rate radium equivalent" (Ra-eq.DR) to cancel the effect of source decay on the variation of dose rate values. RESULTS: Both techniques showed very similar geometry, tandem-ovoid loading ratio, and Ra-eq.DR at A(rev) and A(or). However, the best reproducibility of dose rate was achieved with A(or). Also, time variable-dose rate definition affected the difference between the 2 treatments modalities. CONCLUSIONS: Clear definition of both the point of dose prescription and the time variable used for dose rate calculation are important for brachytherapy for cancer of the cervix.


Assuntos
Braquiterapia/métodos , Dosagem Radioterapêutica , Neoplasias do Colo do Útero/radioterapia , Feminino , Humanos , Reprodutibilidade dos Testes