Upgrade of the JET far infrared interferometer diagnostic.
Boboc, A; Gil, C; Pastor, P; Spuig, P; Edlington, T; Dorling, S.
Affiliation
  • Boboc A; EURATOM/CCFE Association, Culham Science Centre, Abingdon, United Kingdom. Alexandru.Boboc@ccfe.ac.uk
Rev Sci Instrum ; 83(10): 10E341, 2012 Oct.
Article in En | MEDLINE | ID: mdl-23130803
ABSTRACT
In recent years there has been a major upgrade of the JET far infrared diagnostic system consisting of a new laser system with the wavelength at 118.8 µm at and more advanced processing electronics for phase counting. This provides a second colour measurement of the electron plasma density on the vertical system. Due to the shorter wavelength, the plasma induced laser beam refraction is reduced by a factor of three alleviating density errors caused by loss of signal (so-called "fringe jumps" [A. Murari et al., Rev. Sci. Instrum. 77, 073505 (2006)]), in particular during high performance plasmas experiments in JET.

Full text: 1 Collection: 01-internacional Database: MEDLINE Type of study: Diagnostic_studies Language: En Journal: Rev Sci Instrum Year: 2012 Document type: Article