Motion interplay as a function of patient parameters and spot size in spot scanning proton therapy for lung cancer.
Grassberger, Clemens; Dowdell, Stephen; Lomax, Antony; Sharp, Greg; Shackleford, James; Choi, Noah; Willers, Henning; Paganetti, Harald.
Affiliation
  • Grassberger C; Department of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, Massachusetts, USA. Grassberger.Clemens@mgh.harvard.edu
Int J Radiat Oncol Biol Phys ; 86(2): 380-6, 2013 Jun 01.
Article in En | MEDLINE | ID: mdl-23462423
ABSTRACT

PURPOSE:

To quantify the impact of respiratory motion on the treatment of lung tumors with spot scanning proton therapy. METHODS AND MATERIALS Four-dimensional Monte Carlo simulations were used to assess the interplay effect, which results from relative motion of the tumor and the proton beam, on the dose distribution in the patient. Ten patients with varying tumor sizes (2.6-82.3 cc) and motion amplitudes (3-30 mm) were included in the study. We investigated the impact of the spot size, which varies between proton facilities, and studied single fractions and conventionally fractionated treatments. The following metrics were used in the

analysis:

minimum/maximum/mean dose, target dose homogeneity, and 2-year local control rate (2y-LC).

RESULTS:

Respiratory motion reduces the target dose homogeneity, with the largest effects observed for the highest motion amplitudes. Smaller spot sizes (σ ≈ 3 mm) are inherently more sensitive to motion, decreasing target dose homogeneity on average by a factor 2.8 compared with a larger spot size (σ ≈ 13 mm). Using a smaller spot size to treat a tumor with 30-mm motion amplitude reduces the minimum dose to 44.7% of the prescribed dose, decreasing modeled 2y-LC from 87.0% to 2.7%, assuming a single fraction. Conventional fractionation partly mitigates this reduction, yielding a 2y-LC of 71.6%. For the large spot size, conventional fractionation increases target dose homogeneity and prevents a deterioration of 2y-LC for all patients. No correlation with tumor volume is observed. The effect on the normal lung dose distribution is minimal observed changes in mean lung dose and lung V20 are <0.6 Gy(RBE) and <1.7%, respectively.

CONCLUSIONS:

For the patients in this study, 2y-LC could be preserved in the presence of interplay using a large spot size and conventional fractionation. For treatments using smaller spot sizes and/or in the delivery of single fractions, interplay effects can lead to significant deterioration of the dose distribution and lower 2y-LC.
Subject(s)

Full text: 1 Collection: 01-internacional Health context: 1_ASSA2030 Database: MEDLINE Main subject: Respiration / Proton Therapy / Lung Neoplasms / Movement Type of study: Health_economic_evaluation / Observational_studies / Prognostic_studies Limits: Humans Language: En Journal: Int J Radiat Oncol Biol Phys Year: 2013 Document type: Article

Full text: 1 Collection: 01-internacional Health context: 1_ASSA2030 Database: MEDLINE Main subject: Respiration / Proton Therapy / Lung Neoplasms / Movement Type of study: Health_economic_evaluation / Observational_studies / Prognostic_studies Limits: Humans Language: En Journal: Int J Radiat Oncol Biol Phys Year: 2013 Document type: Article