Your browser doesn't support javascript.
loading
Impact of closure devices on vascular complication and mortality rates in TAVI procedures.
Dimitriadis, Zisis; Scholtz, Werner; Börgermann, Jochen; Wiemer, Marcus; Piper, Cornelia; Vlachojannis, Mario; Gummert, Jan; Horstkotte, Dieter; Ensminger, Stephan; Faber, Lothar; Scholtz, Smita.
Affiliation
  • Dimitriadis Z; Clinic for Cardiology, Herz- und Diabeteszentrum Nordrhein-Westfalen, Ruhr-Universität Bochum, Bad Oeynhausen, Germany. Electronic address: andrea.grasmueck@unimedizin-mainz.de.
  • Scholtz W; Clinic for Cardiology, Herz- und Diabeteszentrum Nordrhein-Westfalen, Ruhr-Universität Bochum, Bad Oeynhausen, Germany.
  • Börgermann J; Clinic for Thoracic and Cardiovascular Surgery, Herz- und Diabeteszentrum Nordrhein-Westfalen, Ruhr-Universität Bochum, Bad Oeynhausen, Germany.
  • Wiemer M; Clinic for Cardiology, Herz- und Diabeteszentrum Nordrhein-Westfalen, Ruhr-Universität Bochum, Bad Oeynhausen, Germany.
  • Piper C; Clinic for Cardiology, Herz- und Diabeteszentrum Nordrhein-Westfalen, Ruhr-Universität Bochum, Bad Oeynhausen, Germany.
  • Vlachojannis M; Clinic for Cardiology, Herz- und Diabeteszentrum Nordrhein-Westfalen, Ruhr-Universität Bochum, Bad Oeynhausen, Germany.
  • Gummert J; Clinic for Thoracic and Cardiovascular Surgery, Herz- und Diabeteszentrum Nordrhein-Westfalen, Ruhr-Universität Bochum, Bad Oeynhausen, Germany.
  • Horstkotte D; Clinic for Cardiology, Herz- und Diabeteszentrum Nordrhein-Westfalen, Ruhr-Universität Bochum, Bad Oeynhausen, Germany.
  • Ensminger S; Clinic for Thoracic and Cardiovascular Surgery, Herz- und Diabeteszentrum Nordrhein-Westfalen, Ruhr-Universität Bochum, Bad Oeynhausen, Germany.
  • Faber L; Clinic for Cardiology, Herz- und Diabeteszentrum Nordrhein-Westfalen, Ruhr-Universität Bochum, Bad Oeynhausen, Germany.
  • Scholtz S; Clinic for Cardiology, Herz- und Diabeteszentrum Nordrhein-Westfalen, Ruhr-Universität Bochum, Bad Oeynhausen, Germany.
Int J Cardiol ; 241: 133-137, 2017 Aug 15.
Article in En | MEDLINE | ID: mdl-28153535
ABSTRACT

BACKGROUND:

Currently two closure devices are available for the vascular access in TAVI procedures. Their impact on vascular complications and mid-term mortality is yet unknown.

METHODS:

Between 2009 and 2014, 398 patients underwent TAVI TF procedures in which two different closure devices were used, Prostar® XL (n=215) and Perclose-Proglide® (n=183). In the cases with Prostar we used one device and in cases with Perclose-Proglide, two devices. The two groups were compared with respect to the criteria of the Valve Academic Research Consortium (VARC) II. The mean follow-up period was 679.7±481.8 (727) days.

RESULTS:

There were no significant differences in the baseline characteristics of both patient groups. In the Prostar® group, complications were more frequent (26.6% vs. 12.6%, p=0.005); in detail, these were bleeding (14.9% vs. 7.1%, [p]=0.02), suture rupture (4.7% vs. 1.3%, p=0.04), and pseudoaneurysms (10.2% vs. 1.2%, p<0.001). Multivariate regression analysis revealed two predictors for vascular complications female sex (OR 2.3; 95% CI 1.3-3.8, p=0.002) and closure devices (OR 0.5; 95% CI 0.3-0.8, p=0.007) in favour of Proglide®. There was no significant difference in 30-day mortality (Prostar 5.6±1.6% vs. Proglide 4.9±1.6%). However, Kaplan-Meier survival analysis showed a significantly higher survival rate over the entire follow-up period for the Proglide® group (p=0.03).

CONCLUSION:

Vascular complications occurred more often in the Prostar® group. Although 30-day mortality showed no significant difference between the groups, the mortality over complete follow-up was significantly lower in the Proglide® group.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Postoperative Complications / Equipment Design / Transcatheter Aortic Valve Replacement / Vascular Closure Devices Type of study: Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limits: Aged / Aged80 / Female / Humans / Male Language: En Journal: Int J Cardiol Year: 2017 Document type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Postoperative Complications / Equipment Design / Transcatheter Aortic Valve Replacement / Vascular Closure Devices Type of study: Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limits: Aged / Aged80 / Female / Humans / Male Language: En Journal: Int J Cardiol Year: 2017 Document type: Article