Determining the Orientation of Directional Deep Brain Stimulation Electrodes Using 3D Rotational Fluoroscopy.
Reinacher, P C; Krüger, M T; Coenen, V A; Shah, M; Roelz, R; Jenkner, C; Egger, K.
Affiliation
  • Reinacher PC; From the Departments of Stereotactic and Functional Neurosurgery (P.C.R., V.A.C.) peter.reinacher@uniklinik-freiburg.de.
  • Krüger MT; Neurosurgery (M.T.K., M.S., R.R.).
  • Coenen VA; From the Departments of Stereotactic and Functional Neurosurgery (P.C.R., V.A.C.).
  • Shah M; Neurosurgery (M.T.K., M.S., R.R.).
  • Roelz R; Neurosurgery (M.T.K., M.S., R.R.).
  • Jenkner C; Clinical Trial Unit (C.J.), Freiburg University Medical Center, Freiburg, Germany.
  • Egger K; Neuroradiology (K.E.).
AJNR Am J Neuroradiol ; 38(6): 1111-1116, 2017 Jun.
Article in En | MEDLINE | ID: mdl-28385887
ABSTRACT
BACKGROUND AND

PURPOSE:

New deep brain stimulation leads with electrode contacts that are split along their circumference allow steering of the electrical field in a predefined direction. However, imaging-assisted directional stimulation requires detailed knowledge of the exact orientation of the electrode array. The purpose of this study was to evaluate whether this information can be obtained by rotational 3D fluoroscopy. MATERIALS AND

METHODS:

Two directional leads were inserted into a 3D-printed plaster skull filled with gelatin. The torsion of the lead tip versus the lead at the burr-hole level was investigated. Then, 3 blinded raters evaluated 12 3D fluoroscopies with random lead orientations. They determined the lead orientation considering the x-ray marker only and considering the overlap of the gaps between the contact segments. Intraclass correlation coefficients and an extended version of the Bland-Altman plot were used to determine interrater reliability and agreement of the measurements of the different raters.

RESULTS:

Electrode torsion of up to 35° could be demonstrated. Evaluation of the lead rotation considering the x-ray marker only revealed limits of agreement of ±9.37° and an intraclass correlation coefficient of 0.9975. In addition, taking into account the lines resulting from overlapping of the gaps between the electrode segments, the limits of agreement to the mean were ±2.44° and an intraclass correlation coefficient of 0.9998.

CONCLUSIONS:

In directional deep brain stimulation systems, rotational 3D fluoroscopy combined with the described evaluation method allows for determining the exact orientation of the leads, enabling the full potential of imaging-assisted personalized programming.