Quantification of carotid plaque composition with a multi-contrast atherosclerosis characterization (MATCH) MRI sequence.
Kassem, Mohamed; Nies, Kelly P H; Boswijk, Ellen; van der Pol, Jochem; Aizaz, Mueez; Gijbels, Marion J J; Li, Debiao; Bucerius, Jan; Mess, Werner H; Wildberger, Joachim E; van Oostenbrugge, Robert J; Moonen, Rik P M; Fan, Zhaoyang; Kooi, M Eline.
Affiliation
  • Kassem M; Cardiovascular Research Institute Maastricht (CARIM), Maastricht University Medical Center, Maastricht, Netherlands.
  • Nies KPH; Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, Netherlands.
  • Boswijk E; Cardiovascular Research Institute Maastricht (CARIM), Maastricht University Medical Center, Maastricht, Netherlands.
  • van der Pol J; Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, Netherlands.
  • Aizaz M; Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, Netherlands.
  • Gijbels MJJ; Department of Rehabilitation Medicine, Amsterdam University Medical Center, Location VUmc, Amsterdam, Netherlands.
  • Li D; Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, Netherlands.
  • Bucerius J; Cardiovascular Research Institute Maastricht (CARIM), Maastricht University Medical Center, Maastricht, Netherlands.
  • Mess WH; Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, Netherlands.
  • Wildberger JE; Department of Pathology, Cardiovascular Research Institute Maastricht, Maastricht University Medical Center, Maastricht, Netherlands.
  • van Oostenbrugge RJ; Department of Medical Biochemistry, Experimental Vascular Biology, Amsterdam Cardiovascular Sciences, Amsterdam Infection and Immunity, Amsterdam UMC, Amsterdam, Netherlands.
  • Moonen RPM; Biomedical Imaging Research Institute, Cedars-Sinai Medical Center, Los Angeles, CA, United States.
  • Fan Z; Department of Nuclear Medicine, Georg-August University Göttingen, Universitätsmedizin Göttingen, Göttingen, Germany.
  • Kooi ME; Cardiovascular Research Institute Maastricht (CARIM), Maastricht University Medical Center, Maastricht, Netherlands.
Front Cardiovasc Med ; 10: 1227495, 2023.
Article in En | MEDLINE | ID: mdl-37680565
ABSTRACT
Background and

purpose:

Carotid atherosclerotic plaques with a large lipid-rich necrotic core (LRNC), intraplaque hemorrhage (IPH), and a thin or ruptured fibrous cap are associated with increased stroke risk. Multi-sequence MRI can be used to quantify carotid atherosclerotic plaque composition. Yet, its clinical implementation is hampered by long scan times and image misregistration. Multi-contrast atherosclerosis characterization (MATCH) overcomes these limitations. This study aims to compare the quantification of plaque composition with MATCH and multi-sequence MRI.

Methods:

MATCH and multi-sequence MRI were used to image 54 carotid arteries of 27 symptomatic patients with ≥2 mm carotid plaque on a 3.0 T MRI scanner. The following sequence parameters for MATCH were used repetition time/echo time (TR/TE), 10.1/4.35 ms; field of view, 160 mm × 160 mm × 2 mm; matrix size, 256 × 256; acquired in-plane resolution, 0.63 mm2× 0.63 mm2; number of slices, 18; and flip angles, 8°, 5°, and 10°. Multi-sequence MRI (black-blood pre- and post-contrast T1-weighted, time of flight, and magnetization prepared rapid acquisition gradient echo; acquired in-plane resolution 0.63 mm2 × 0.63 mm2) was acquired according to consensus recommendations, and image quality was scored (5-point scale). The interobserver agreement in plaque composition quantification was assessed by the intraclass correlation coefficient (ICC). The sensitivity and specificity of MATCH in identifying plaque composition were calculated using multi-sequence MRI as a reference standard.

Results:

A significantly lower image quality of MATCH compared to that of multi-sequence MRI was observed (p < 0.05). The scan time for MATCH was shorter (7 vs. 40 min). Interobserver agreement in quantifying plaque composition on MATCH images was good to excellent (ICC ≥ 0.77) except for the total volume of calcifications and fibrous tissue that showed moderate agreement (ICC ≥ 0.61). The sensitivity and specificity of detecting plaque components on MATCH were ≥89% and ≥91% for IPH, ≥81% and 85% for LRNC, and ≥71% and ≥32% for calcifications, respectively. Overall, good-to-excellent agreement (ICC ≥ 0.76) of quantifying plaque components on MATCH with multi-sequence MRI as the reference standard was observed except for calcifications (ICC = 0.37-0.38) and fibrous tissue (ICC = 0.59-0.70). Discussion and

conclusion:

MATCH images can be used to quantify plaque components such as LRNC and IPH but not for calcifications. Although MATCH images showed a lower mean image quality score, short scan time and inherent co-registration are significant advantages.