Factors associated with damages and repair costs of reusable flexible bronchoscopes in pediatrics in China.
Hu, Ruixue; Chen, Yanhua; Jiang, Sixin; Hu, Ting; Huang, Juanli; Zhang, Lu; Pan, Wei; Yi, Liangying; Huang, Yan.
Affiliation
  • Hu R; Department of Sterile Processing Nursing, West China Second University Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Chen Y; Key Laboratory of Birth Defects and Related Diseases of Women and Children (Sichuan University), Ministry of Education, Chengdu, Sichuan, China.
  • Jiang S; Department of Sterile Processing Nursing, West China Second University Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Hu T; Key Laboratory of Birth Defects and Related Diseases of Women and Children (Sichuan University), Ministry of Education, Chengdu, Sichuan, China.
  • Huang J; Department of Sterile Processing Nursing, West China Second University Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Zhang L; Key Laboratory of Birth Defects and Related Diseases of Women and Children (Sichuan University), Ministry of Education, Chengdu, Sichuan, China.
  • Pan W; Department of Sterile Processing Nursing, West China Second University Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Yi L; Key Laboratory of Birth Defects and Related Diseases of Women and Children (Sichuan University), Ministry of Education, Chengdu, Sichuan, China.
  • Huang Y; Department of Sterile Processing Nursing, West China Second University Hospital, Sichuan University, Chengdu, Sichuan, China.
BMC Pediatr ; 24(1): 228, 2024 Apr 01.
Article in En | MEDLINE | ID: mdl-38561723
ABSTRACT

BACKGROUND:

Frequent repairs of pediatric flexible bronchoscopes can lead to a huge financial burden for the hospital. This study aimed to investigate the common causes of the failures in pediatric flexible bronchoscopes and propose the measures to prevent the failures.

METHODS:

This was a retrospective study. We collected repair information of the pediatric flexible bronchoscopes reprocessed in the Department of Sterile Processing at a hospital between September 1, 2018 and September 1, 2022 in order to investigate the causes and possible factors associated with the failures in pediatric flexible bronchoscopes.

RESULTS:

The Department of Sterile Processing staff reprocessed the pediatric flexible bronchoscopes 4280 times. A total of 29 failures were identified. The failure rate was 0.678%. The average repair cost was USD7246.60. The common failures in the pediatric flexible bronchoscopes included dim video image, black dots, improper video image display or no image during angulation adjustment, and pressure marks in the insertion tube. The failure rates in flexible electronic bronchoscopes and small-diameter flexible bronchoscopes were 65.5% and 93.1%, respectively. The failure rate in the pediatric flexible bronchoscopes reprocessed by the staff members with less work experience was 75.9%.

CONCLUSION:

The failure rate in the pediatric flexible bronchoscopes was not high but the repair costs were extremely high. The types and size of the flexible bronchoscopes and work experience of the staff members responsible for bronchoscope reprocessing were the possible factors associated with the failure rate in the pediatric flexible bronchoscopes. It is advisable to further optimize the central workflow and management mode for reprocessing the pediatric flexible bronchoscopes, thereby extending their useful life and reducing costs.