Your browser doesn't support javascript.
loading
A channel-homogenized micro gas chromatography column for light alkane separation with uniform HKUST-1 stationary phase.
Ma, Shaojie; Zhu, Yuchen; Chen, Boxin; Zhao, Bin; Feng, Fei.
Affiliation
  • Ma S; State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China; Center of Materials Science and Optoelectronics Engineering, University of Chinese Academy of Sciences, Beijing 100049, China.
  • Zhu Y; State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China; Center of Materials Science and Optoelectronics Engineering, University of Chinese Academy of Sciences, Beijing 100049, China.
  • Chen B; State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China; Center of Materials Science and Optoelectronics Engineering, University of Chinese Academy of Sciences, Beijing 100049, China.
  • Zhao B; State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China.
  • Feng F; State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China; Center of Materials Science and Optoelectronics Engineering, University of Chinese Academy of Sciences, Beijing 100049, China. Electronic
Anal Chim Acta ; 1307: 342619, 2024 Jun 08.
Article in En | MEDLINE | ID: mdl-38719412
ABSTRACT

BACKGROUND:

The micro gas chromatography column (µGCC) is one of the key components of the miniaturized gas chromatography system. However, light alkanes are difficult to be separated by a micro gas chromatography column, especially for methane and ethane, because the length of µGCC is limited by the area of a silicon substrate. More importantly, the heterogeneous microchannel surface formed by silicon glass bonding causes uneven stationary phase coating and the forces between the untreated microchannel surfaces and the stationary phase materials are weak, which will prevent the improvement of separation performance.

RESULTS:

In this paper, a micro gas chromatography column (µGCC) with uniform HKUST-1 stationary phase is reported. Significantly, an alumina film prepared by the atomic layer deposition (ALD) technique is used to homogenize the heterogeneous microchannels. The alumina is a hydrophilic material and the alumina made by the ALD technique is uniform. The forces between hydrophilic alumina film and HKUST-1 are strong, which can greatly improve the coating uniformity of the hydrophilic stationary phase HKUST-1. The test results show that the µGCC could baseline separate the light alkane mixtures (CH4, C2H6, C3H8, and C4H10) at the high testing temperature of 120 °C. The maximum resolution of the difficult-separated methane and ethane reached 19.2, which is 108 % higher than the µGCC using the same stationary phase without homogenizing the microchannel inner surface.

SIGNIFICANCE:

The µGCC uses ALD alumina film to homogenize the microchannel inner surface; meanwhile, hydrophilic ALD alumina has a strong electrostatic attraction with the hydrophilic stationary phase HKUST-1. Homogeneous microchannel surface and strong electrostatic attraction are favorable to obtain uniform stationary phase which greatly improves the separation performance, resulting in a large resolution for methane and ethane. The µGCC has broad application prospects in light alkane separation.