Fat quantification in dual-layer detector spectral CT: How to handle iron overload, varying tube voltage and radiation dose Indices.
Molwitz, Isabel; Campbell, Graeme Michael; Knopp, Tobias; Schubert, Niklas; Erley, Jennifer; Löser, Anastassia; Adam, Gerhard; Yamamura, Jin; Fischer, Roland; Ozga, Ann-Kathrin; Szwargulski, Patryk.
Affiliation
  • Molwitz I; Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
  • Campbell GM; Philips GmbH Market DACH, Hamburg, Germany.
  • Knopp T; Institute for Biomedical Imaging, Technical University, Hamburg, Germany.
  • Schubert N; Section for Biomedical Imaging, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
  • Erley J; Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
  • Löser A; Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
  • Adam G; Department of Radiotherapy, University Hospital Schleswig-Holstein Campus Lübeck, Lübeck, Germany.
  • Yamamura J; Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
  • Fischer R; Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
  • Ozga AK; Evidia Group, Berlin, Germany.
  • Szwargulski P; Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
PLoS One ; 19(5): e0302863, 2024.
Article in En | MEDLINE | ID: mdl-38781228
ABSTRACT

OBJECTIVES:

Opposed to other spectral CT techniques, fat quantification in dual-layer detector CT (dlCT) has only recently been developed. The impact of concomitant iron overload and dlCT-specific protocol settings such as the dose right index (DRI), a measure of image noise and tube current, on dlCT fat quantification was unclear. Further, spectral information became newly available <120 kV. Therefore, this study's objective was to evaluate the impact of iron, changing tube voltage, and DRI on dlCT fat quantification. MATERIAL AND

METHODS:

Phantoms with 0 and 8mg/cm3 iron; 0 and 5mg/cm3 iodine; 0, 10, 20, 35, 50, and 100% fat and liver equivalent, respectively, were scanned with a dlCT (CT7500, Philips, the Netherlands) at 100kV/20DRI, 120kV/20DRI, 140kV/20DRI, and at 120kV/16DRI, 120kV/24DRI. Material decomposition was done for fat, liver, and iodine (A1); for fat, liver, and iron (A2); and for fat, liver, and combined reference values of iodine and iron (A3). All scans were analyzed with reference values from 120kV/20DRI. For statistics, the intraclass correlation coefficient (ICC) and Bland-Altman analyses were used.

RESULTS:

In phantoms with iron and iodine, results were best for A3 with a mean deviation to phantom fat of 1.3±2.6% (ICC 0.999 [95%-confidence interval 0.996-1]). The standard approach A1 yielded a deviation of -2.5±3.0% (0.998[0.994-0.999]), A2 of 6.1±4.8% (0.991[0.974-0.997]). With A3 and changing tube voltage, the maximal difference between quantified fat and the phantom ground truth occurred at 100kV with 4.6±2.1%. Differences between scans were largest between 100kV and 140kV (2.0%[-7.1-11.2]). The maximal difference of changing DRI occurred between 16 and 24 DRI with 0.4%[-2.2-3.0].

CONCLUSION:

For dlCT fat quantification in the presence of iron, material decomposition with combined reference values for iodine and iron delivers the most accurate results. Tube voltage-specific calibration of reference values is advisable while the impact of the DRI on dlCT fat quantification is neglectable.