Your browser doesn't support javascript.
loading
Measurement of left ventricular volume with admittance incorporated onto percutaneous ventricular assist device.
Diaz Sanmartin, Luis A; Gruslova, Aleksandra B; Nolen, Drew R; Feldman, Marc D; Valvano, Jonathan W.
Affiliation
  • Diaz Sanmartin LA; Department of Medicine, Division of Cardiology, University of Texas Health Science Center at San Antonio, San Antonio, TX, USA.
  • Gruslova AB; Department of Medicine, Division of Cardiology, University of Texas Health Science Center at San Antonio, San Antonio, TX, USA.
  • Nolen DR; Department of Medicine, Division of Cardiology, University of Texas Health Science Center at San Antonio, San Antonio, TX, USA.
  • Feldman MD; Department of Medicine, Division of Cardiology, University of Texas Health Science Center at San Antonio, San Antonio, TX, USA.
  • Valvano JW; Department of Electrical Engineering, University of Texas at Austin, Austin, TX, USA. valvano@mail.utexas.edu.
Med Biol Eng Comput ; 2024 Jul 15.
Article in En | MEDLINE | ID: mdl-39004652
ABSTRACT
Percutaneous ventricular assist devices (pVADs) incorporated with admittance electrodes have been validated in animal studies for accurate instantaneous volumetric measurements. Since miniaturization of the pVAD profile is a priority to reduce vascular complications in patients, our study aimed to validate admittance measurements using three electrodes instead of the standard four. Complex admittance was measured between an electrode pair and a pVAD metallic blood-intake tip, both with finite element analysis and on the benchtop. The catheter and electrode arrays were first simulated inside prolate ellipsoid models of the left ventricle (LV) demonstrating current flow throughout all parts of the LV as well as minimal influence of off-center catheter placement in the recorded signal. Admittance measurements were validated in 3D-printed models of healthy and dilated hearts (100-400 mL end-diastolic volumes). Minimal interference between a pVAD motor and the current signal of our admittance system was demonstrated. A modified Wei's equation focused on three electrodes was developed to be compatible with reduced profile pVADs occurring clinically, incorporated with admittance electrodes and wires. The modified equation was compared against Wei's original equation showing improved accuracy of calculated volumes. Reducing electrode footprint can simplify the incorporation of Admittance technology on any pVAD, allowing for instantaneous recognition of native heart recovery and assistance with pVAD weaning.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Med Biol Eng Comput Year: 2024 Document type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Med Biol Eng Comput Year: 2024 Document type: Article