Urinálise: comparação entre microscopia óptica e citometria de fluxo / Urinalysis: comparison between microscopic and flow cytometry analysis
Bottini, P. V; Garlipp, C. R.
Afiliação
  • Bottini, P. V; Universidade Estadual de Campinas. Hospital de Clínicas. Divisão de Patologia Clínica. Campinas. BR
  • Garlipp, C. R; UNICAMP. Faculdade de Ciências Médicas. Departamento de Patologia Clínica. Campinas. BR
J. bras. patol. med. lab ; 42(3): 157-162, jun. 2006. graf, tab
Article em Pt | LILACS | ID: lil-452996
Biblioteca responsável: BR14.1
RESUMO
INTRODUÇÃO E

OBJETIVO:

O exame de urina é um procedimento de alta demanda, trabalhoso e pouco padronizado. Este estudo teve por objetivo avaliar o desempenho de um citômetro de fluxo na realização do exame de urina de rotina. CASUíSTICA E

MÉTODOS:

Foram analisadas 1.140 amostras de urina através de microscopia óptica comum e de citometria de fluxo (UF-100/SYSMEX). A precisão foi estabelecida com a análise de quatro amostras de urina (20 replicações cada). O cálculo da reprodutibilidade foi realizado a partir de 30 determinações de dois controles comerciais em dias consecutivos.

RESULTADOS:

As contagens de hemácias e leucócitos mostraram concordância de 91 por cento e 93 por cento, respectivamente. Cilindros, células e bactérias mostraram sobreposição dos valores fornecidos pelo UF-100 quando comparados com os relatados na análise microscópica. A precisão do UF-100 variou de 4 por cento a 155 por cento, com reprodutibilidade de 3 por cento e 25 por cento, dependendo do parâmetro avaliado.

CONCLUSÃO:

O equipamento UF-100/SYSMEX demonstra boa precisão, reprodutibilidade e concordância com a microscopia óptica. A utilização da citometria de fluxo implica numa maior agilização e padronização da rotina, bem como em uma nova maneira de reportar e interpretar o exame de urina de rotina.
ABSTRACT

INTRODUCTION:

Urinalysis is a high demand procedure, with large amount of manual labor and poorly standardized. The purpose of this investigation was to analyze the performance of an automated system based on flow cytometry for routine urinalysis. MATERIAL AND

METHODS:

We analyzed 1,140 urine samples by light field microscopy and by flow cytometry (UF-100/SYSMEX). For the precision study of the UF-100, we calculated the within-run and between-run coefficients of variation using two different levels of commercial controls and four different urine samples.

RESULTS:

Erythrocytes and leukocytes counts by the two methods showed an agreement of 91 percent and 93 percent, respectively. Casts, epithelial cells and bacteria counts by the UF-100 showed a significant overlap when compared to microscopic analysis. Intra assay precision (within-run) ranged from 4 percent to 155 percent and interassay precision (between-run) varied from 3 percent to 25 percent, depending on the considered parameter.

CONCLUSION:

Flow cytometry is a precise and reproducible technique, with a strong correlation with the results obtained by microscopic analysis. Flow cytometry allows a better workflow and a new manner of reporting and interpreting routine urinalysis.
Assuntos
Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Urina / Reprodutibilidade dos Testes / Urinálise / Citometria de Fluxo / Microscopia Tipo de estudo: Guideline Idioma: Pt Revista: J. bras. patol. med. lab Ano de publicação: 2006 Tipo de documento: Article
Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Urina / Reprodutibilidade dos Testes / Urinálise / Citometria de Fluxo / Microscopia Tipo de estudo: Guideline Idioma: Pt Revista: J. bras. patol. med. lab Ano de publicação: 2006 Tipo de documento: Article