ABSTRACT
Procedures for implantation of mechanical aortic valves have to consider eccentric flow in the aortic root. We describe how to optimize orientation of tilting disc (Medtronic Hall) and bileaflet (St. Jude Medical) valves. In tilting disc valves, an asymmetric design faces an asymmetric flow. Hemodynamic performance of this valve type, regarding turbulence and pressure gradients, is close to normal physiology and superior to the bileaflet valve design. This difference is more pronounced the smaller the valve size.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Próteses Valvulares Cardíacas / Implante de Prótese de Valva Cardíaca / Hemodinâmica Tipo de estudo: Clinical_trials / Observational_studies Limite: Humans Idioma: En Revista: Ann Thorac Surg Ano de publicação: 1999 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Próteses Valvulares Cardíacas / Implante de Prótese de Valva Cardíaca / Hemodinâmica Tipo de estudo: Clinical_trials / Observational_studies Limite: Humans Idioma: En Revista: Ann Thorac Surg Ano de publicação: 1999 Tipo de documento: Article