Your browser doesn't support javascript.
loading
Critical analysis of dual-chamber implantable cardioverter-defibrillator arrhythmia detection : results and technical considerations.
Wilkoff, B L; Kühlkamp, V; Volosin, K; Ellenbogen, K; Waldecker, B; Kacet, S; Gillberg, J M; DeSouza, C M.
Afiliação
  • Wilkoff BL; Dept of Cardiology, Cleveland Clinic Foundation, Cleveland, Ohio, USA.
Circulation ; 103(3): 381-6, 2001 Jan 23.
Article em En | MEDLINE | ID: mdl-11157689
ABSTRACT

BACKGROUND:

One of the perceived benefits of dual-chamber implantable cardioverter-defibrillators (ICDs) is the reduction in inappropriate therapy due to new detection algorithms. It was the purpose of the present investigation to propose methods to minimize bias during such comparisons and to report the arrhythmia detection clinical results of the PR Logic dual-chamber detection algorithm in the GEM DR ICD in the context of these methods. METHODS AND

RESULTS:

Between November 1997 and October 1998, 933 patients received the GEM DR ICD in this prospective multicenter study. A total of 4856 sustained arrhythmia episodes (n=311) with stored electrogram and marker channel were classified by the investigators; 3488 episodes (n=232) were ventricular tachycardia (VT)/ventricular fibrillation (VF), and 1368 episodes (n=149) were supraventricular tachycardia (SVT). The overall detection results were corrected for multiple episodes within a patient with the generalized estimating equations (GEE) method with an exchangeable correlation structure between episodes. The relative sensitivity for detection of sustained VT and/or VF was 100.0% (3488 of 3488, n=232; 95% CI 98.3% to 100%), the VT/VF positive predictivity was 88.4% uncorrected (3488 of 3945, n=278) and 78.1% corrected (95% CI 73.3% to 82.3%) with the GEE method, and the SVT positive predictivity was 100.0% (911 of 911, n=101; 95% CI 96% to 100%).

CONCLUSIONS:

A structured approach to analysis limits the bias inherent in the evaluation of tachycardia discrimination algorithms through the use of relative VT/VF sensitivity, VT/VF positive predictivity, and SVT positive predictivity along with corrections for multiple tachycardia episodes in a single patient.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Desfibriladores Implantáveis Tipo de estudo: Clinical_trials / Diagnostic_studies / Evaluation_studies / Observational_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Circulation Ano de publicação: 2001 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Desfibriladores Implantáveis Tipo de estudo: Clinical_trials / Diagnostic_studies / Evaluation_studies / Observational_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Circulation Ano de publicação: 2001 Tipo de documento: Article