Sensitive determination of diquat by a kinetic method using the stopped-flow mixing technique.
Pérez-Ruiz, T; Martínez-Lozano, C; Tomás, V; Fenoll, J.
Afiliação
  • Pérez-Ruiz T; Department of Analytical Chemistry, Faculty of Chemistry, University of Murcia, 30071 Murcia, Spain. tpr@fcu.um.es
Analyst ; 125(12): 2372-5, 2000 Dec.
Article em En | MEDLINE | ID: mdl-11219084
ABSTRACT
A simple, sensitive, selective, fast and inexpensive assay for the determination of diquat is proposed. The method is based on the reduction of the herbicide to a strongly fluorescent monocation radical with sodium dithionite. The initial rate of this reaction is directly proportional to the diquat concentration. The stopped-flow mixing technique was used because the kinetic data can be obtained in only 7 s, meaning that the method can be automated. The calibration graph is linear over the range 5-500 ng ml-1 and the precision (RSD) is close to 1.2%. The applicability of the method was demonstrated by determining the herbicide in different kinds of samples.
Assuntos
Buscar no Google
Coleções: 01-internacional Contexto em Saúde: 2_ODS3 Base de dados: MEDLINE Assunto principal: Diquat / Poluentes Ambientais / Herbicidas Tipo de estudo: Diagnostic_studies Idioma: En Revista: Analyst Ano de publicação: 2000 Tipo de documento: Article