Novel magnetic technology for intraoperative intracranial frameless navigation: in vivo and in vitro results.
Zaaroor, M; Bejerano, Y; Weinfeld, Z; Ben-Haim, S.
Afiliação
  • Zaaroor M; Department of Neurosurgery, Rambam Medical Center, Haifa, Israel. mzaaroor@hotmail.com
Neurosurgery ; 48(5): 1100-7; discussion 1107-8, 2001 May.
Article em En | MEDLINE | ID: mdl-11334277
ABSTRACT

OBJECTIVE:

To characterize the accuracy of the Magellan electromagnetic navigation system (Biosense Webster, Tirat HaCarmel, Israel) and to demonstrate the feasibility of its use in image-guided neurosurgical applications. DESCRIPTION OF INSTRUMENTATION The Magellan system was developed to provide real-time tracking of the distal tips of flexible catheters, steerable endoscopes, and other surgical instruments, using ultra-low electromagnetic fields and a novel miniature position sensor for image-correlated intraoperative navigation and mapping applications.

METHODS:

An image registration procedure was performed, and static and qualitative accuracies were assessed in a series of phantom, animal, and human neurosurgical studies. EXPERIENCE AND

RESULTS:

During the human study phase, an accuracy error of up to 5 mm was deemed acceptable. Results demonstrated that this degree of accuracy was maintained throughout all procedures. All anatomic landmarks were reached with precision and were accurately viewed on the display screen. Navigation that relied on the system was also successful. No interference with operating room equipment was noted. The accuracy of the system was maintained during regular surgical procedures, using standard surgical tools.

CONCLUSION:

The system provides precise lesion localization without limiting the line of vision, the mobility of the surgeon, or the flexibility of instruments. Electromagnetic navigation promises new advances in neuronavigation and frameless stereotactic surgery.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Terapia Assistida por Computador / Encéfalo / Tecnologia Radiológica / Magnetismo / Neurocirurgia Tipo de estudo: Evaluation_studies / Qualitative_research Limite: Animals / Humans Idioma: En Revista: Neurosurgery Ano de publicação: 2001 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Terapia Assistida por Computador / Encéfalo / Tecnologia Radiológica / Magnetismo / Neurocirurgia Tipo de estudo: Evaluation_studies / Qualitative_research Limite: Animals / Humans Idioma: En Revista: Neurosurgery Ano de publicação: 2001 Tipo de documento: Article