Your browser doesn't support javascript.
loading
Catastrophic failure of the Elite Plus total hip replacement, with a Hylamer acetabulum and Zirconia ceramic femoral head.
Norton, M R; Yarlagadda, R; Anderson, G H.
Afiliação
  • Norton MR; Department of Trauma and Orthopaedic Surgery, Derriford Hospital, Plymouth, England.
J Bone Joint Surg Br ; 84(5): 631-5, 2002 Jul.
Article em En | MEDLINE | ID: mdl-12188475
ABSTRACT
We report catastrophic early failure of a cemented total hip replacement comprising a modular femoral component with a Zirconia ceramic head and an acetabular component of cross-linked ultra-high molecular-weight polyethylene (Hylamer). Between 1995 and 1999 we implanted 29 hips in 26 patients with a mean age of 49.2 years. Survivorship analysis in this group revealed a failure rate of 67.6% at five years. All hips which failed did so because of aseptic loosening with progressive osteolysis or radiolucencies. We therefore recommend early and regular review of all patients with this combination of implants and early revision surgery in order to avoid massive bone loss.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Falha de Prótese / Prótese de Quadril Tipo de estudo: Observational_studies Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: J Bone Joint Surg Br Ano de publicação: 2002 Tipo de documento: Article