Reduction of early elastic recoil by cutting balloon angioplasty as compared to conventional balloon angioplasty.
Kawaguchi, Katsuhiro; Kondo, Taizo; Shumiya, Takanori; Uchikawa, Tomohiro; Suzuki, Tomomichi; Awaji, Yoshifumi; Mochizuki, Mitsuhiro.
Afiliação
  • Kawaguchi K; Komaki City Hospital, Department of Cardiology, 1-20, Jobushi, 4858520 AICHI, Japan. kawacci@aurora.dti.ne.jp
J Invasive Cardiol ; 14(9): 515-9, 2002 Sep.
Article em En | MEDLINE | ID: mdl-12205350
ABSTRACT

BACKGROUND:

Restenosis after successful balloon angioplasty remains problematic. Early elastic response after angioplasty is significant when considering the possible development of restenosis. The purpose of this study was to compare early elastic recoil within 10 minutes after successful percutaneous transluminal coronary angioplasty and early lumen loss at 24 hours after angioplasty in a cutting balloon group and a conventional balloon group.

METHODS:

Extent of early elastic recoil was quantitatively measured as the difference of mean balloon diameter at maximal inflation pressure and minimal luminal diameter after angioplasty in 82 cutting balloon-treated lesions and 51 conventional balloon-treated lesions.

RESULTS:

Reference diameter and balloon/artery ratio were similar between the cutting balloon and conventional balloon groups (2.89 0.47 mm vs. 2.88 0.60 mm; 1.19 0.11 vs. 1.19 0.13, respectively). Early elastic recoil after angioplasty was significantly smaller in the cutting balloon than the conventional balloon group (0.96 0.40 mm vs. 1.12 0.37 mm, respectively; p = 0.04). Also, the mean amount of lumen loss from 10 minutes after angioplasty to 24 hours after was significantly smaller in the cutting balloon than the conventional balloon group (0.08 0.28 mm vs. 0.20 0.33 mm, respectively; p = 0.02).

CONCLUSION:

There is significantly less early elastic recoil in the cutting balloon angioplasty than in the conventional balloon angioplasty group. The efficacy of cutting balloon continues 24 hours after angioplasty.
Assuntos
Buscar no Google
Coleções: 01-internacional Contexto em Saúde: 2_ODS3 Base de dados: MEDLINE Assunto principal: Angioplastia com Balão Tipo de estudo: Observational_studies / Prognostic_studies Limite: Aged / Female / Humans / Male / Middle aged País/Região como assunto: Asia Idioma: En Revista: J Invasive Cardiol Ano de publicação: 2002 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Contexto em Saúde: 2_ODS3 Base de dados: MEDLINE Assunto principal: Angioplastia com Balão Tipo de estudo: Observational_studies / Prognostic_studies Limite: Aged / Female / Humans / Male / Middle aged País/Região como assunto: Asia Idioma: En Revista: J Invasive Cardiol Ano de publicação: 2002 Tipo de documento: Article