Anterior occiput to axis screw fixation: part II: a biomechanical comparison with posterior fixation techniques.
Dvorak, Marcel F S; Sekeramayi, Floyd; Zhu, Qingan; Hoekema, Joel; Fisher, Charles; Boyd, Mike; Goertzen, Darrell J; Oxland, Thomas R.
Afiliação
  • Dvorak MF; Division of Spine, Department of Orthopaedics, University of British Columbia and the Combined Neurosurgical and Orthopaedic Spine Program, Vancouver Hospital and Health Sciences Centre, Canada. mdvorak@vanhosp.bc.ca
Spine (Phila Pa 1976) ; 28(3): 239-45, 2003 Feb 01.
Article em En | MEDLINE | ID: mdl-12567024
ABSTRACT
STUDY

DESIGN:

This biomechanical study used flexibility testing on fresh-frozen human cadaveric specimens (occiput to C3) and compared the range of motion and neutral zone for three occipitocervical fixation techniques.

OBJECTIVES:

To contrast the stabilization provided by a new technique of anterior occipitocervical screw fixation with two other commonly used posterior occipitocervical fixation techniques. SUMMARY OF BACKGROUND DATA There are no published reports describing this novel technique of anterior occipitocervical screw fixation.

METHODS:

Six human occipitocervical spine specimens were mounted in a custom-designed, spine-testing machine that applied a pure moment in flexion-extension, lateral bending, and axial rotation. The specimens were tested intact, after an odontoid osteotomy with capsular injury, and after each of three fixation

methods:

posterior wiring, posterior plate fixation with C1-C2 transarticular screws, and finally with anterior occipitocervical screws. Intervertebral motion was measured with an optoelectronic measurement system, and the range of motion and neutral zone were the kinematic variables measured and used for analysis.

RESULTS:

In flexion and extension testing, the posterior plate with transarticular screws provided greater stabilization than posterior wiring or anterior occipitocervical screws. In lateral bending and rotation, the anterior screws were similarly effective to the posterior plate, both of which were more effective than posterior wiring.

CONCLUSION:

The anterior screw fixation technique was as effective as a posterior plate with transarticular screws in stabilizing between the occiput and C2 in axial rotation and lateral bending. In extension and flexion, the anterior screw technique was not as effective as a posterior plate with transarticular screws in providing stability.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fusão Vertebral / Vértebra Cervical Áxis / Parafusos Ósseos / Osso Occipital Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: Spine (Phila Pa 1976) Ano de publicação: 2003 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fusão Vertebral / Vértebra Cervical Áxis / Parafusos Ósseos / Osso Occipital Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: Spine (Phila Pa 1976) Ano de publicação: 2003 Tipo de documento: Article