A rapid method for electron beam energy check.
Pisciotta, V J; Wu, X; Fiedler, J A; Serago, C F; Calish-Watzich, M L; Schwade, J G; Houdek, P V.
Afiliação
  • Pisciotta VJ; Department of Radiation Oncology, University of Miami School of Medicine, Florida 33101.
Med Phys ; 19(6): 1451-3, 1992.
Article em En | MEDLINE | ID: mdl-1461209
ABSTRACT
Assessment of electron beam energy and its long term stability is part of standard quality assurance practice in radiation oncology. Conventional depth-ionization or depth-film density measurements are time consuming both in terms of data acquisition and analysis. A procedure is described utilizing ionization measurements at two energy specific depths. It is based on a linear relationship between electron beam energy and its practical range. Energy shifts within the range covered by the two measurement depths are easily resolved. Within a range of +/- 0.50 MeV (+/- 1.30 MeV) around the established mean incident energy of 5.48 MeV (20.39 MeV), the method accuracy is better than 0.10 MeV.