Pegylated interferon-alpha2a kinetics during experimental haemodialysis: impact of permeability and pore size of dialysers.
Barril, G; Quiroga, J A; Sanz, P; Rodrìguez-Salvanés, F; Selgas, R; Carreño, V.
Afiliação
  • Barril G; Department of Nephrology, Hospital U. de la Princesa, Madrid, Spain. gbarril@wanadoo.es
Aliment Pharmacol Ther ; 20(1): 37-44, 2004 Jul 01.
Article em En | MEDLINE | ID: mdl-15225169
BACKGROUND: Therapeutics in end-stage renal disease (ESRD) patients undergoing haemodialysis (HD) has to consider potential drug clearance during the dialysis procedure. Pegylated interferon-alpha (PEG-IFN-alpha), a middle-size protein drug active against viral hepatitis, allows convenient once-weekly dosing due to prolonged plasma half-life. AIM: To investigate the impact of permeability and dialyser pore size on PEG-IFN-alpha blood levels during experimental HD. METHODS: Polymethylmetacrylate (PMMA) membrane 1.6 m2 dialysers with three different permeabilities/pore sizes were selected. RESULTS: A 40 kDa PEG-IFN-alpha2a (PEGASYS) was not cleared (< 5%) through low-flux/small pore size (25 A;B3A) and high-flux/middle-large pore size (60 A;BKP) dialysers, and was partially (approximately 15%) through intermediate permeability/large pore size (100 A;BKF) dialysers. In contrast, unmodified 17 kDa IFN-alpha2a(Roferon-A) was removed (65%-95%) through BKP or BKF, but not B3A, PMMA dialysers. Moreover, 12 kDa PEG-IFN-alpha2b(PegIntron) was cleared (40%-80%) through PMMA dialysers with pore sizes > or = 60 angstroms. When B3A or BKP were replaced every hour PEG-IFN-alpha2a plasma levels remained constant throughout three experimental-HD-sessions, but PEG-IFN-alpha2b was cleared partially every BKP replacement. Porosity differ among high-flux dialysers. Neither PEG-IFN-alpha2a nor PEG-IFN-alpha2b were removed after three HD sessions through (27/31/33 A) pore size polysulphone dialysers. Although PEG-IFN-alpha2a was not cleared through middle pore-size (43 A/AN69ST) polyacrylonitrile dialyser, PEG-IFN-alpha2b was partially removed. CONCLUSIONS: The pharmacokinetics of Peg-IFN-alpha may vary in a patient on dialysis.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Antivirais / Polietilenoglicóis / Diálise Renal / Interferon-alfa Limite: Humans Idioma: En Revista: Aliment Pharmacol Ther Ano de publicação: 2004 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Antivirais / Polietilenoglicóis / Diálise Renal / Interferon-alfa Limite: Humans Idioma: En Revista: Aliment Pharmacol Ther Ano de publicação: 2004 Tipo de documento: Article