Extraction of optical scattering parameters and attenuation compensation in optical coherence tomography images of multilayered tissue structures.
Thrane, Lars; Frosz, Michael H; Jørgensen, Thomas M; Tycho, Andreas; Yura, Harold T; Andersen, Peter E.
Afiliação
  • Thrane L; Optics and Plasma Research Department, Risø National Laboratory, Roskilde, Denmark. lars.thrane@risoe.dk
Opt Lett ; 29(14): 1641-3, 2004 Jul 15.
Article em En | MEDLINE | ID: mdl-15309845
A recently developed analytical optical coherence tomography (OCT) model [Thrane et al., J. Opt. Soc. Am. A 17, 484 (2000)] allows the extraction of optical scattering parameters from OCT images, thereby permitting attenuation compensation in those images. By expanding this theoretical model, we have developed a new method for extracting optical scattering parameters from multilayered tissue structures in vivo. To verify this, we used a Monte Carlo (MC) OCT model as a numerical phantom to simulate the OCT signal for heterogeneous multilayered tissue. Excellent agreement between the extracted values of the optical scattering properties of the different layers and the corresponding input reference values of the MC simulation was obtained, which demonstrates the feasibility of the method for in vivo applications. This is to our knowledge the first time such verification has been obtained, and the results hold promise for expanding the functional imaging capabilities of OCT.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Tecido Conjuntivo / Tomografia de Coerência Óptica Tipo de estudo: Evaluation_studies Idioma: En Revista: Opt Lett Ano de publicação: 2004 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Tecido Conjuntivo / Tomografia de Coerência Óptica Tipo de estudo: Evaluation_studies Idioma: En Revista: Opt Lett Ano de publicação: 2004 Tipo de documento: Article