Your browser doesn't support javascript.
loading
Evaluation of Elispot assays: influence of method and operator on variability of results.
Janetzki, S; Schaed, S; Blachere, N E B; Ben-Porat, L; Houghton, A N; Panageas, K S.
Afiliação
  • Janetzki S; ZellNet Consulting, Inc., 555 North Avenue, Suite 25-S, Fort Lee NJ 07024, USA. sylvia@zellnet.com
J Immunol Methods ; 291(1-2): 175-83, 2004 Aug.
Article em En | MEDLINE | ID: mdl-15345315
ABSTRACT
In this study, a comprehensive comparative analysis of different evaluation methods of Elispot plates was performed. Three investigators using three different evaluation approaches read 50 randomly selected wells at three different time points. The methods were (1) manual evaluation using a stereomicroscope, (2) automated evaluation using an image analysis reader system with reading parameters established by each investigator, and (3) automated evaluation using a reader system with preset reading parameters using assay-specific controls. We demonstrate that manual evaluation had the highest variability both within the same method and when comparing all methods, followed by automated evaluation with investigator-dependent parameters. The variability was low only when all investigators used the same parameters established using assay-specific controls. This variability was independent of operator or spot number per well. Based on this study, recommendations for standardization and validation procedures of Elispot assay performance and evaluation procedures are presented.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas Imunoenzimáticas Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: J Immunol Methods Ano de publicação: 2004 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas Imunoenzimáticas Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: J Immunol Methods Ano de publicação: 2004 Tipo de documento: Article