Development and validation of a liquid chromatography-tandem mass spectrometric method for the determination of alpha-methyltyrosine in human plasma.
J Chromatogr B Analyt Technol Biomed Life Sci
; 810(2): 229-34, 2004 Oct 25.
Article
em En
| MEDLINE
| ID: mdl-15380719
A sensitive and selective LC-MS-MS method for the isolation and quantification of alpha-methyltyrosine (AMT) from human plasma is described. The method employs a simple protein precipitation using zinc sulfate and sodium hydroxide. This precipitation procedure produced samples with high aqueous content that could be directly injected into a LC-MS-MS system without compromising reverse-phase chromatographic performance. Chromatographic separation was performed on a MetaChem MonoChrom C(18) column (2.0 mm x 50 mm; 5 microm) at a flow rate of 1 mL/min. Compounds were eluted using a gradient mixture of water-acetic acid (100:0.1, v/v) and acetonitrile-acetic acid (100:0.1, v/v). The structural analog alpha-hydroxymethyltyrosine was used as the internal standard. Mass spectrometric detection was carried out with a triple quadrupole mass spectrometer. The method was validated and used to determine human plasma AMT concentrations, and has been implemented to derive pharmacokinetic parameters.