Close-geometry efficiency calibration of p-type HPGe detectors with a Cs-134 point source.
Appl Radiat Isot
; 64(10-11): 1303-6, 2006.
Article
em En
| MEDLINE
| ID: mdl-16549352
When close-geometry detector calibration is required in gamma-ray spectrometry, single-line emitters are usually used in order to avoid true coincidence summing effects. We managed to overcome this limitation by developing a method for the determination of the efficiency of p-type HPGe detectors in close-geometry with a calibrated Cs-134 point source. No separate determination of coincidence summing correction factors is required and a single measurement furnishes the full-energy-peak efficiencies in the 475-1365 keV energy range.