Close-geometry efficiency calibration of p-type HPGe detectors with a Cs-134 point source.
DeFelice, P; Fazio, A; Vidmar, T; Korun, M.
Afiliação
  • DeFelice P; EC Joint Research Centre Ispra, I-21020 Italy.
Appl Radiat Isot ; 64(10-11): 1303-6, 2006.
Article em En | MEDLINE | ID: mdl-16549352
When close-geometry detector calibration is required in gamma-ray spectrometry, single-line emitters are usually used in order to avoid true coincidence summing effects. We managed to overcome this limitation by developing a method for the determination of the efficiency of p-type HPGe detectors in close-geometry with a calibrated Cs-134 point source. No separate determination of coincidence summing correction factors is required and a single measurement furnishes the full-energy-peak efficiencies in the 475-1365 keV energy range.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Transdutores / Algoritmos / Radioisótopos de Césio Tipo de estudo: Diagnostic_studies / Evaluation_studies / Guideline / Prognostic_studies País/Região como assunto: Europa Idioma: En Revista: Appl Radiat Isot Ano de publicação: 2006 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Transdutores / Algoritmos / Radioisótopos de Césio Tipo de estudo: Diagnostic_studies / Evaluation_studies / Guideline / Prognostic_studies País/Região como assunto: Europa Idioma: En Revista: Appl Radiat Isot Ano de publicação: 2006 Tipo de documento: Article