Determination of oxytetracycline levels in rainbow trout serum on a biphenyl column using high-performance liquid chromatography.
Miller, Ron A; Reimschuessel, Renate; Carson, Mary C.
Afiliação
  • Miller RA; U.S. Food and Drug Administration, Center for Veterinary Medicine, Office of Research, 8401 Muirkirk Road, Laurel, MD 20708, USA. Ron.Miller@fda.hhs.gov
J Chromatogr B Analyt Technol Biomed Life Sci ; 852(1-2): 655-8, 2007 Jun 01.
Article em En | MEDLINE | ID: mdl-17301001
ABSTRACT
We developed a simple and sensitive high-performance liquid chromatography method on a biphenyl column to determine oxytetracycline (OTC) levels in rainbow trout serum. The assay used deproteination, filtration, and subsequent separation on a reverse-phase biphenyl column, with UV detection at 355 nm. OTC (7.8-7.9 min) was completely resolved from structurally similar riboflavin (10.4-10.5 min), a common feed supplement. Estimated limits of detection and quantitation of OTC were 0.01 and 0.04 microg/mL, respectively. The average recovery for OTC was 102% with a R.S.D. of 8.34%. Calibration standards were linear from 0.01 to 10 microg/mL.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxitetraciclina / Cromatografia Líquida de Alta Pressão Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci Ano de publicação: 2007 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxitetraciclina / Cromatografia Líquida de Alta Pressão Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci Ano de publicação: 2007 Tipo de documento: Article