Determination of oxytetracycline levels in rainbow trout serum on a biphenyl column using high-performance liquid chromatography.
J Chromatogr B Analyt Technol Biomed Life Sci
; 852(1-2): 655-8, 2007 Jun 01.
Article
em En
| MEDLINE
| ID: mdl-17301001
ABSTRACT
We developed a simple and sensitive high-performance liquid chromatography method on a biphenyl column to determine oxytetracycline (OTC) levels in rainbow trout serum. The assay used deproteination, filtration, and subsequent separation on a reverse-phase biphenyl column, with UV detection at 355 nm. OTC (7.8-7.9 min) was completely resolved from structurally similar riboflavin (10.4-10.5 min), a common feed supplement. Estimated limits of detection and quantitation of OTC were 0.01 and 0.04 microg/mL, respectively. The average recovery for OTC was 102% with a R.S.D. of 8.34%. Calibration standards were linear from 0.01 to 10 microg/mL.