Evaluation of controlled-drift detectors in X-ray spectroscopic imaging applications.
Castoldi, Andrea; Guazzoni, Chiara; Ozkan, Cigdem; Vedani, Giorgio; Hartmann, Robert; Bjeoumikhov, Aniouar.
Afiliação
  • Castoldi A; Politecnico di Milano, Dip. Elettronica e Informazione, P.za Leonardo da Vinci 32, 20133 Milano, Italy. andrea.castoldi@polimi.it
Microsc Microanal ; 15(3): 231-6, 2009 Jun.
Article em En | MEDLINE | ID: mdl-19460179
ABSTRACT
A detector that looks promising for advanced imaging modalities--such as X-ray absorption contrast imaging, X-ray fluorescence imaging, and diffraction-enhanced imaging--is the controlled-drift detector (CDD). The CDD is a novel two-dimensional X-ray imager with energy resolving capability of spectroscopic quality. It is built on a fully depleted silicon wafer and features fast readout while being operated at or near room temperature. The use of CDDs in the aforementioned applications allows translating these techniques from synchrotron-based experiments to laboratory-size experiments using polychromatic X-ray generators. We have built a dedicated and versatile detection module based on a 36 mm2 CDD chip featuring pixels of 180 x 180 microm 2, and we evaluated the system performance in different X-ray imaging applications both with synchrotron-based experiments and in the laboratory environment.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Microsc Microanal Ano de publicação: 2009 Tipo de documento: Article