Your browser doesn't support javascript.
loading
Imaging of soft tissues adjacent to orthopedic hardware: comparison of 3-T and 1.5-T MRI.
Farrelly, Cormac; Davarpanah, Amir; Brennan, Stephen A; Brennan, Stephen; Sampson, Mathew; Eustace, Stephen J.
Afiliação
  • Farrelly C; Department of Radiology, Cappagh National Orthopaedic Hospital, Finglas, Dublin 11, Republic of Ireland. farrellycormac@gmail.com
AJR Am J Roentgenol ; 194(1): W60-4, 2010 Jan.
Article em En | MEDLINE | ID: mdl-20028892
ABSTRACT

OBJECTIVE:

The purpose of this study was to compare metal artifact reduction techniques at 1.5-T and 3-T MRI. MATERIALS AND

METHODS:

A titanium plate with steel screws was placed in a freshly harvested pig leg. The leg was imaged with 1.5-T and 3-T MRI. A T2-weighted turbo spin-echo sequence was used with echo-train lengths of 8, 16, 32, and 64 and a constant readout bandwidth of 31.2 kHz. The images were compared qualitatively, and the optimal echo-train length was selected. Images were acquired at the optimal echo-train length with four different readout bandwidths. Artifact was measured quantitatively, and image quality was ranked qualitatively. The qualitatively best image acquired at 1.5 T was compared with the qualitatively highest-ranked image acquired at 3 T.

RESULTS:

At both 1.5 T and 3 T, optimal images of equal quality were produced at echo-train lengths of 8 and 16. At higher readout bandwidths, there was quantitatively less artifact. The qualitatively best images were acquired at a readout bandwidth of 31.2 kHz at 1.5 T and 62.5 kHz at 3 T (Cronbach's alpha=1.00). The optimal image at 3 T was qualitatively superior to that at 1.5 T.

CONCLUSION:

Optimizing image acquisition parameters in this phantom model resulted in similar quantitative susceptibility artifact at 3 T and 1.5 T and better qualitative images at 3 T than at 1.5 T.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Placas Ósseas / Parafusos Ósseos / Imageamento por Ressonância Magnética / Aumento da Imagem / Artefatos Tipo de estudo: Qualitative_research Limite: Animals Idioma: En Revista: AJR Am J Roentgenol Ano de publicação: 2010 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Placas Ósseas / Parafusos Ósseos / Imageamento por Ressonância Magnética / Aumento da Imagem / Artefatos Tipo de estudo: Qualitative_research Limite: Animals Idioma: En Revista: AJR Am J Roentgenol Ano de publicação: 2010 Tipo de documento: Article