Your browser doesn't support javascript.
loading
Computed tomography commissioning programmes: how to obtain a reliable MTF with an automatic approach?
Miéville, F; Beaumont, S; Torfeh, T; Gudinchet, F; Verdun, F R.
Afiliação
  • Miéville F; Institute of Radiation Physics IRA, University Hospital Center and University of Lausanne, Grand Pré 1, 1007 Lausanne, Switzerland.
Radiat Prot Dosimetry ; 139(1-3): 443-8, 2010.
Article em En | MEDLINE | ID: mdl-20167797
The purpose of this study was to assess the spatial resolution of a computed tomography (CT) scanner with an automatic approach developed for routine quality controls when varying CT parameters. The methods available to assess the modulation transfer functions (MTF) with the automatic approach were Droege's and the bead point source (BPS) methods. These MTFs were compared with presampled ones obtained using Boone's method. The results show that Droege's method is not accurate in the low-frequency range, whereas the BPS method is highly sensitive to image noise. While both methods are well adapted to routine stability controls, it was shown that they are not able to provide absolute measurements. On the other hand, Boone's method, which is robust with respect to aliasing, more resilient to noise and provides absolute measurements, satisfies the commissioning requirements perfectly. Thus, Boone's method combined with a modified Catphan 600 phantom could be a good solution to assess CT spatial resolution in the different CT planes.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Algoritmos / Reconhecimento Automatizado de Padrão / Interpretação de Imagem Radiográfica Assistida por Computador / Intensificação de Imagem Radiográfica / Tomografia Computadorizada por Raios X Tipo de estudo: Diagnostic_studies Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2010 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Algoritmos / Reconhecimento Automatizado de Padrão / Interpretação de Imagem Radiográfica Assistida por Computador / Intensificação de Imagem Radiográfica / Tomografia Computadorizada por Raios X Tipo de estudo: Diagnostic_studies Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2010 Tipo de documento: Article